What is Wood’s metal?
Fusible plus that melts at elevated temperatures.
When does the oxygen pressure failure device activate?
What is the hypoxia prevention safety device?
Annular space
The area between the widest area of the indicator field and side wall of flowmeter.
Thorpe tube
- Is the beginning of the low pressure system.
which position of the O2 flowmeter minimizes the risk of a hypoxia mixture in the event of a flowmeter leak?
O2 flowmeter should be positioned all the way to the right (closest to the common gas outlet)
Reynolds Number
Predicts flow rate through a tube and determine when the gas flow is laminar or turbulent
Reynolds number = DensityxDiameterxvelocty/viscosity
Flowmeter FiO2 calculation
FiO2 =
Airflow rate x21)+(O2 flow rate x100
————————————-
Total flow rate
How to calculate tidal volume with fresh gas coupling
Total tidal volume =
Vt set on ventilator + FGF during inspiration - volume lost to compliance
Circuit compliance
Compliance =
Change in volume
————————-
Change in pressure
Variable bypass vaporizer
How many pressure systems does an anesthesia machine have?
-3
What is included in the high pressure system of an anesthesia machine?
What is cylinder supply used for?
- tanks should only be open when checked or when pipeline supply is unavailable
How many psi are in the following compressed gas cylinders?
What is the PISS
What is the cylinder valve?
What is safety relief devises of Cylinder valve?
What is hanger yoke?
What is included in the intermediate pressure system?
What problems can occur with pipeline supply?
-pressure loss or excess pressure
-cross-connection of gas delivery pipeline
-leaks
Theft of nitrous oxide
What do you do if oxygen pipeline failure?
What are the 5 tasks of oxygen in the AGM?
What devices (or techniques) do not rely on wall outlet power?