Gas Cylinder Colors
Empty E-Cylinder Weight
5.8 lbs
Circle System Check
Extends from CGO to the Y-pieceLeak test: occlude y-piece, close APL, pressurize to 30 mmHg using the O2 flush valve
Machine Check Overview (8 Items)
*check analyzer by exposing to room air and reading 21%
Intermediate Pressure System (7 Items)
5 Pathways of Oxygen
Low Pressure Circuit Test
Checks from flow control valves to CGO. Place suction valve on CGO and squeeze till collapsed; must remain collapsed for 10 seconds or more. Also check for low pressure O2 alarm; check individual vaporizers for leaks
Oxygen Failsafe Shut Off Pressure
25-30 psi
Most are Proportioning Type (OFPD); drop N2O based on O2 PRESSURE not FiO2.
Located downstream of N2O source; shuts off or proportionally reduces N2O flow if O2 decreases to prevent hypoxia
PISS system configurations:
Pin Index Safety System
Bobbin Flow Types
Oxygen Cylinder: contents and pressure
625-700L; 2000-2000 psi
First Stage Pressure Regulator: 1. O2 2. N2O Second Stage Pressure Regulator: 1. O2 2. N20
First Stage: 02: 2000 -> 45 N2O: 745 -> 45 Second Stage: O2: 45 -> 14 N2O: 45 -> 26
*N2O at higher pressure so that it flushes out of the system if there’s a leak
Low Pressure System (6 items)
Pressure slightly above atmospheric
Oxygen Flush Valve flow rate
35-75 L/minCan cause barotrauma; only use during expiration
High Pressure System Components (5 items)
Pressure > 45 psi
Pressurized Air Cylinder: Contents and Pressure
Approx 625L; 1800 psi
Common Vaporizer Problems (7 items)
Describe Minimal Sedation:
Describe Moderate Sedation:
Describe Deep Sedation:
Describe General Sedation:
Anesthesia Workstation Standards overseen by…
American Society for Testing and Materials (ASTM)
Machine Checks Should be done…
2. Circuit check between each patient