ch23 _Technical_Considerations_DECK Flashcards

(192 cards)

1
Q

Which of the following BEST describes digital exposure considerations?
a. Traditional radiographic principles no longer apply
b. Only software determines image quality
c. Traditional parameters still apply in digital imaging
d. Exposure factors are irrelevant

A

c. Traditional parameters still apply in digital imaging

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Which factor is MOST directly related to distance and intensity changes in digital imaging?
a. Inverse square law
b. Grid ratio
c. Filtration
d. Collimation

A

a. Inverse square law

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Which exposure factor improves spatial resolution when decreased?
a. SID
b. OID
c. mAs
d. Filtration

A

b. OID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Which of the following is TRUE regarding focal spot size?
a. Larger focal spot improves resolution
b. Smaller focal spot improves resolution
c. Focal spot size does not affect image quality
d. Focal spot only affects brightness

A

b. Smaller focal spot improves resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Which factor primarily controls beam penetration in digital imaging?
a. mAs
b. SID
c. kVp
d. Grid

A

c. kVp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is the PRIMARY role of mAs in digital imaging?
a. Controls contrast
b. Controls brightness only
c. Controls quantity of radiation (detector exposure)
d. Controls spatial resolution

A

c. Controls quantity of radiation (detector exposure)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Why are high mA and short exposure times preferred?
a. Increase contrast
b. Reduce motion blur
c. Increase noise
d. Reduce penetration

A

b. Reduce motion blur

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What is the MAIN purpose of collimation?
a. Increase brightness
b. Reduce scatter and patient dose
c. Increase magnification
d. Improve focal spot size

A

b. Reduce scatter and patient dose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the PRIMARY benefit of using grids?
a. Increase noise
b. Reduce scatter radiation
c. Increase magnification
d. Decrease resolution

A

b. Reduce scatter radiation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the PURPOSE of filtration in radiography?
a. Increase low-energy photons
b. Remove low-energy photons
c. Increase image noise
d. Increase patient dose

A

b. Remove low-energy photons

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Which patient factor MOST affects exposure selection?
a. Hair color
b. Patient size and condition
c. Age only
d. Gender

A

b. Patient size and condition

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Which of the following improves radiographic spatial resolution in DR?
a. Decreasing SID
b. Increasing SID
c. Increasing OID
d. Increasing filtration

A

b. Increasing SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What happens to spatial resolution when SID is increased?
a. Decreases due to magnification
b. No change
c. Improves due to reduced magnification
d. Increases noise

A

c. Improves due to reduced magnification

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Which focal spot size provides the BEST spatial resolution?
a. Large focal spot
b. Medium focal spot
c. Small focal spot
d. Focal spot size does not matter

A

c. Small focal spot

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Why does a smaller focal spot improve resolution?
a. Increases beam intensity
b. Reduces penumbra
c. Increases scatter
d. Decreases exposure time

A

b. Reduces penumbra

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is the PRIMARY benefit of using high mA with short exposure time?
a. Increases contrast
b. Reduces motion blur
c. Increases magnification
d. Reduces penetration

A

b. Reduces motion blur

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Which of the following combinations BEST improves image sharpness?
a. Low mA, long time
b. High mA, long time
c. High mA, short time
d. Low mA, short time

A

c. High mA, short time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is the effect of motion on radiographic images?
a. Increases contrast
b. Improves resolution
c. Causes image blur
d. Reduces noise

A

c. Causes image blur

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Which factor is MOST responsible for geometric sharpness?
a. kVp
b. mAs
c. Focal spot size
d. Filtration

A

c. Focal spot size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What happens to image sharpness when OID increases?
a. Improves resolution
b. No change
c. Decreases resolution due to magnification
d. Reduces noise

A

c. Decreases resolution due to magnification

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

In digital imaging, what is the relationship between kVp and contrast?
a. kVp directly controls contrast like film-screen
b. kVp has no effect on the image
c. kVp only controls brightness
d. kVp does not control contrast in the traditional sense

A

d. kVp does not control contrast in the traditional sense

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

In digital radiography, what primarily determines image brightness?
a. mAs
b. kVp
c. Distance
d. Post-processing (LUT)

A

d. Post-processing (LUT)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Why are traditional mAs and image density relationships no longer valid in digital imaging?
a. mAs no longer affects exposure
b. Density is controlled by software processing
c. mAs only affects contrast now
d. mAs is replaced by kVp

A

b. Density is controlled by software processing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

What does the detector convert x-ray exposure into in digital imaging?
a. Heat
b. Light only
c. Electronic signal
d. Magnetic field

A

c. Electronic signal

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
What is the MOST important factor for digital image quality? a. kVp selection only b. mAs selection only c. Total exposure to the detector d. Focal spot size
c. Total exposure to the detector
26
Which statement is TRUE about mAs in digital imaging? a. It controls brightness directly b. It has no role in image quality c. It controls the quantity of photons reaching the detector d. It only affects contrast
c. It controls the quantity of photons reaching the detector
27
What happens when exposure to the detector is too low? a. Image becomes too bright b. Image becomes noisy (quantum mottle) c. Image contrast increases d. Detector saturation occurs
b. Image becomes noisy (quantum mottle)
28
What happens when exposure to the detector is excessively high? a. Increased noise b. Reduced signal c. No effect at all d. Increased patient dose without improving diagnostic quality
d. Increased patient dose without improving diagnostic quality
29
Which statement BEST describes digital contrast? a. Controlled primarily by kVp b. Controlled primarily by mAs c. Controlled primarily by post-processing algorithms d. Controlled by SID
c. Controlled primarily by post-processing algorithms
30
In digital imaging, how does kVp selection typically compare to analog imaging? a. Lower kVp is used b. Higher kVp is used c. kVp is eliminated d. kVp only affects brightness
b. Higher kVp is used
31
What is the primary benefit of using higher kVp in digital imaging? a. Reduces ESE b. Increases image noise c. Decreases penetration d. Increases patient dose
a. Reduces entrance skin exposure (ESE)
32
What does the 15% rule state? a. Increase kVp by 15% → double mAs b. Increase kVp by 15% → halve mAs c. Decrease kVp by 15% → halve mAs d. kVp and mAs are unrelated
b. Increase kVp by 15% → halve mAs
33
What happens when mAs is too low in digital imaging? a. Image becomes too bright b. Contrast increases c. Quantum mottle (noise) increases d. Detector saturation occurs
c. Quantum mottle (noise) increases
34
What is the MAIN cause of quantum mottle? a. Excessive kVp b. Insufficient photon quantity c. Excessive filtration d. High SID
b. Insufficient photon quantity
35
What is the MOST important factor in digital image quality? a. kVp selection b. Focal spot size c. SID d. Total exposure to the detector
d. Total exposure to the detector
36
Why can digital imaging tolerate higher kVp techniques? a. kVp controls brightness b. Software compensates for contrast changes c. kVp has no effect on the image d. mAs becomes irrelevant
b. Software compensates for contrast changes
37
What happens if exposure to the detector is adequate but kVp is high? a. Image will always be too dark b. Image will have no signal c. Image can still appear acceptable due to processing d. Image will always have high contrast
c. Image can still appear acceptable due to processing
38
What is the effect of image noise on diagnostic quality? a. Improves diagnostic acuity b. Has no effect on diagnosis c. Impairs diagnostic acuity d. Only affects brightness
c. Impairs diagnostic acuity
39
How is quantum noise primarily controlled in digital imaging? a. Adjusting SID b. Modifying mAs c. Changing focal spot size d. Using filtration
b. Modifying mAs
40
What happens to image noise when mAs is increased? a. Noise increases b. Noise decreases c. No change in noise d. Contrast decreases
b. Noise decreases
41
What is the result of using excessively high mAs? a. Photon starvation b. Increased noise c. Reduced signal d. Detector element (DEL) saturation
d. Detector element (DEL) saturation
42
What artifact can result from detector saturation? a. Grid cutoff b. Motion blur c. Data drop d. Fogging
c. Data drop
43
What does 'data drop' refer to? a. Loss of image brightness b. Loss of signal due to underexposure c. Detector overwhelmed and unable to record data d. Increase in scatter radiation
c. Detector overwhelmed and unable to record data
44
What is the purpose of exposure indicators (EI)? a. Measure patient size b. Control image contrast c. Assess detector exposure and image quality d. Adjust focal spot size
c. Assess detector exposure and image quality
45
What is the relationship between mAs and image signal? a. mAs decreases signal b. mAs has no effect on signal c. mAs only affects contrast d. Higher mAs increases signal
d. Higher mAs increases signal
46
Why are exposure technique systems still important in digital imaging? a. They control image brightness directly b. They eliminate the need for EI values c. They ensure consistent exposure and image quality d. They replace post-processing
c. They ensure consistent exposure and image quality
47
What is image quality primarily referenced to in digital imaging? a. kVp b. Exposure Indicator (EI) c. SID d. Focal spot size
b. Exposure Indicator (EI)
48
Why can exposure indicator (EI) values be confusing in clinical practice? a. EI only measures contrast b. EI values change with SID c. EI is not used in digital imaging d. Different manufacturers use different EI scales
d. Different manufacturers use different EI scales
49
What must departments provide to ensure proper exposure technique? a. Only kVp charts b. Only mAs charts c. Exposure technique charts and target EI values d. Only post-processing software
c. Exposure technique charts and target EI values
50
What do target EI values represent? a. Maximum allowable patient dose b. Standard of acceptable detector exposure c. Image brightness level d. Focal spot size standard
b. Standard of acceptable detector exposure
51
Who originally developed exposure technique systems in the 1920s? a. Arthur W. Fuchs b. Wilhelm Roentgen c. Ed. C. Jerman d. Marie Curie
c. Ed. C. Jerman
52
Who refined exposure technique systems into a scientific method? a. Ed. C. Jerman b. Arthur W. Fuchs c. Thomas Edison d. Isaac Newton
b. Arthur W. Fuchs
53
What are the two main exposure technique systems? a. High kVp and low kVp systems b. Analog and digital systems c. Fixed kVp/variable mAs and variable kVp/fixed mAs d. Manual and automatic systems
c. Fixed kVp/variable mAs and variable kVp/fixed mAs
54
Which exposure technique system is BEST suited for digital imaging? a. Variable kVp, fixed mAs b. Fixed kVp, variable mAs c. Fixed mAs only d. Variable SID system
b. Fixed kVp, variable mAs
55
What is the defining feature of a fixed kVp system? a. kVp remains constant while mAs is adjusted b. mAs remains constant while kVp is adjusted c. Both kVp and mAs remain constant d. Neither kVp nor mAs are controlled
a. kVp remains constant while mAs is adjusted
56
What must be established first in a fixed kVp system? a. Patient weight b. Optimal kVp for the body part c. SID d. Grid ratio
b. Optimal kVp for the body part
57
What is the purpose of selecting an optimal kVp? a. Control image brightness b. Reduce exposure time c. Ensure desired image contrast d. Eliminate scatter
c. Ensure desired image contrast
58
In a fixed kVp system, what parameter is adjusted for patient thickness? a. kVp b. SID c. Focal spot size d. mAs
d. mAs
59
What is the minimum percentage change in mAs considered significant? a. 10% b. 15% c. 30% d. 50%
c. 30%
60
In a fixed kVp system, how is mAs adjusted for every 5 cm increase in patient thickness? a. Decrease mAs by half b. Double mAs c. Increase kVp d. No change
b. Double mAs
61
In a fixed kVp system, how is mAs adjusted for every 5 cm decrease in thickness? a. Double mAs b. Increase kVp c. Reduce SID d. Halve mAs
d. Halve mAs
62
Why are mAs changes less than 30% typically avoided? a. They significantly increase dose b. They do not produce noticeable changes in exposure c. They affect contrast too much d. They increase noise dramatically
b. They do not produce noticeable changes in exposure
63
What is the defining feature of a variable kVp system? a. kVp is adjusted based on patient thickness b. mAs remains constant at all times c. Both kVp and mAs remain constant d. Only SID is adjusted
a. kVp is adjusted based on patient thickness
64
What is the typical threshold kVp used in a variable kVp system? a. 15 kVp b. 30 kVp c. 60 kVp d. 90 kVp
b. 30 kVp
65
In a variable kVp system, how often is kVp adjusted based on thickness? a. Every 1 cm b. Every 5 cm c. Every 2 cm d. Every 10 cm
c. Every 2 cm
66
What is the correct formula used to calculate new kVp in this system? a. (2 × part cm) + 30 kVp b. (5 × part cm) + 15 kVp c. (2 × part cm) − 30 kVp d. (30 × part cm) + 2 kVp
a. (2 × part cm) + 30 kVp
67
Which exposure rule is applied when using the variable kVp system? a. Inverse square law b. 30% rule c. 5 cm rule d. 15% rule
d. 15% rule
68
Why is a threshold kVp required in this system? a. To determine starting contrast level b. To eliminate scatter radiation c. To reduce exposure time d. To fix mAs values
a. To determine starting contrast level
69
If a body part measures 10 cm, what is the calculated kVp using this system? a. 20 kVp b. 30 kVp c. 50 kVp d. 70 kVp
c. 50 kVp
70
What happens to kVp as part thickness increases in this system? a. It decreases b. It stays constant c. It increases d. It becomes irrelevant
c. It increases
71
What is the first step in establishing a technique system? a. Place chart into clinical trial b. Test exposures using a phantom c. Create exposure chart d. Re-examine chart
b. Test exposures using a phantom
72
Who is consulted to determine acceptability of test exposures? a. Patient b. Radiologic technologist and/or radiologist c. Nurse d. Physicist only
b. Radiologic technologist and/or radiologist
73
What is created after initial phantom testing? a. Final exposure chart b. Clinical trial report c. Theoretical exposure chart d. Exposure index log
c. Theoretical exposure chart
74
What is the purpose of using a range of large and small sizes when creating a chart? a. Reduce radiation dose b. Improve positioning accuracy c. Standardize SID d. Account for patient variability
d. Account for patient variability
75
What is done after the theoretical exposure chart is created? a. Re-examine for accuracy b. Adjust exposure index c. Place chart into clinical trial d. Discard incorrect data
c. Place chart into clinical trial
76
What is the purpose of clinical fine tuning? a. Adjust chart using real patient images and measurements b. Eliminate need for exposure charts c. Fix kVp permanently d. Reduce grid usage
a. Adjust chart using real patient images and measurements
77
What is the final step in establishing a technique system? a. Clinical trial b. Phantom testing c. Re-examining and adjusting over time d. Creating theoretical chart
c. Re-examining and adjusting over time
78
What is the primary factor used to judge image acceptability? a. Contrast b. Brightness c. Spatial resolution d. Image noise
d. Image noise
79
What is used as a reference point for evaluating exposures? a. SID b. Exposure Index (EI) c. kVp chart d. Grid ratio
b. Exposure Index (EI)
80
What is the relationship between image noise and exposure? a. Inversely related b. Directly related c. Not related d. Only related in CR
a. Inversely related
81
Why is digital imaging more tolerant of overexposure? a. Increased contrast resolution b. Use of grids c. Digital post-processing software d. Lower kVp settings
c. Digital post-processing software
82
What happens with underexposure in digital imaging? a. Increased exposure latitude b. Decreased exposure latitude c. No change in image quality d. Increased brightness
b. Decreased exposure latitude
83
What is 'dose creep'? a. Decreasing patient dose over time b. Increasing kVp for better contrast c. Gradual increase in exposure due to reliance on post-processing d. Reducing mAs to minimum levels
c. Gradual increase in exposure due to reliance on post-processing
84
What primarily causes dose creep? a. Fear of underexposure and reliance on rescaling b. Incorrect SID c. Low grid ratios d. Poor positioning
a. Fear of underexposure and reliance on rescaling
85
What is the recommended exposure technique approach in digital imaging? a. Low kVp and high mAs b. High kVp and low mAs c. Low kVp and low mAs d. High kVp and high mAs
b. High kVp and low mAs
86
Why is underexposure more problematic than overexposure in digital imaging? a. It increases patient dose b. It reduces contrast c. It causes increased image noise d. It affects positioning
c. It causes increased image noise
87
What does increasing exposure generally do to image noise? a. Increases noise b. Decreases noise c. Has no effect d. Only affects brightness
b. Decreases noise
88
What is ALARA compliance primarily based on in digital radiography? a. kVp selection b. Target Exposure Index (EIₜ) values c. SID adjustments d. Grid usage
b. Target Exposure Index (EIₜ) values
89
What is the purpose of using accurate exposure techniques based on patient measurement? a. Reduce positioning time b. Improve image brightness c. Optimize radiation dose and image quality d. Eliminate need for EI
c. Optimize radiation dose and image quality
90
Who is responsible for establishing target EI values? a. Radiologist only b. Equipment manufacturer c. Individual technologist d. Department
d. Department
91
What is the Deviation Index (DI) based on? a. kVp settings b. Target Exposure Index (EIₜ) c. Patient size d. SID
b. Target Exposure Index (EIₜ)
92
Where can EI and DI values typically be found? a. On the control panel only b. In the PACS image annotation only c. In the DICOM header of the image d. In the exposure chart
c. In the DICOM header of the image
93
What does a Deviation Index (DI) represent? a. Patient positioning accuracy b. Difference between actual EI and target EI c. Image contrast level d. Amount of scatter radiation
b. Difference between actual EI and target EI
94
Why is ALARA important in digital imaging? a. To increase image brightness b. To reduce radiation dose while maintaining image quality c. To eliminate need for technique charts d. To standardize SID
b. To reduce radiation dose while maintaining image quality
95
What is the main role of Exposure Index (EI)? a. Measure image contrast b. Indicate patient dose level at detector c. Control kVp automatically d. Replace mAs settings
b. Indicate patient dose level at detector
96
Who created the Deviation Index (DI)? a. ASRT b. FDA c. AAPM d. ARRT
c. AAPM
97
What does the Deviation Index (DI) establish? a. Patient positioning accuracy b. Range of EI variances based on target EI c. Image contrast level d. Exposure time
b. Range of EI variances based on target EI
98
What does a positive DI value indicate? a. Underexposure b. Proper exposure c. Equipment error d. Overexposure
d. Overexposure
99
What does a negative DI value indicate? a. Overexposure b. Underexposure c. Correct exposure d. High contrast
b. Underexposure
100
What is the approximate DI scale range? a. ±5 b. ±10 c. ±25 d. ±50
c. ±25
101
At what DI value does exposure become a concern for ALARA? a. ±1 b. ±2 c. ±3 or greater d. ±5
c. ±3 or greater
102
What does a DI of 0 represent? a. Severe overexposure b. Severe underexposure c. Target exposure achieved d. Equipment malfunction
c. Target exposure achieved
103
Why was the DI scale standardized? a. To eliminate exposure charts b. To account for manufacturer differences in EI calculations c. To control SID automatically d. To reduce scan time
b. To account for manufacturer differences in EI calculations
104
What are the primary factors that determine image quality in DR? a. kVp and SID only b. Noise, artifacts, resolution, and DQE c. mAs and grid ratio only d. Exposure time and focal spot size
b. Noise, artifacts, resolution, and DQE
105
What does DQE stand for? a. Digital Quality Enhancement b. Detector Quantum Energy c. Dose Quality Evaluation d. Detective Quantum Efficiency
d. Detective Quantum Efficiency
106
Scatter (secondary radiation) production is primarily affected by which factor? a. kVp b. SID c. mAs d. Grid frequency
c. mAs
107
What is photon starvation caused by? a. Overexposure b. Underexposure c. High kVp d. Large SID
b. Underexposure
108
What happens during photon starvation? a. Excess scatter is produced b. Image becomes overly bright c. Detector saturates d. Insufficient photons reach the detector
d. Insufficient photons reach the detector
109
What can extreme overexposure cause in DR systems? a. Increased noise b. Improved contrast c. Data drop due to DEL saturation d. Reduced resolution
c. Data drop due to DEL saturation
110
What is the main consequence of DEL saturation? a. Loss of image data b. Increased photon starvation c. Improved image detail d. Reduced exposure index
a. Loss of image data
111
Which factor is NOT listed as a primary determinant of DR image quality? a. Noise b. Artifacts c. Resolution d. Patient positioning
d. Patient positioning
112
When assessing DR image quality, what is meant by evaluating the image in 'totality'? a. Only checking exposure index b. Evaluating all image quality factors together c. Focusing only on anatomy d. Ignoring artifacts
b. Evaluating all image quality factors together
113
Which of the following is NOT part of assessing DR image quality? a. Collimation b. Motion distortion c. Patient billing information d. Resolution
c. Patient billing information
114
What does quantum noise (mottle) indicate? a. Excessive exposure b. Insufficient photon interaction c. High contrast d. Detector saturation
b. Insufficient photon interaction
115
Why is collimation important when assessing image quality? a. It increases brightness b. It improves patient comfort c. It reduces scatter and improves image quality d. It controls SID
c. It reduces scatter and improves image quality
116
What type of distortion is caused by patient movement? a. Geometric distortion b. Quantum mottle c. Motion distortion d. Contrast distortion
c. Motion distortion
117
What role does EI play in image assessment? a. Determines positioning accuracy b. Serves as a reference for exposure to the detector c. Controls contrast d. Eliminates artifacts
b. Serves as a reference for exposure to the detector
118
When should an image be sent for interpretation? a. When EI is high regardless of appearance b. When image looks good even if EI is off c. When both image appearance and EI are acceptable d. When artifacts are present
c. When both image appearance and EI are acceptable
119
Which factor ensures correct anatomy is included in the image? a. Resolution b. Positioning c. Brightness d. EI value
b. Positioning
120
What is electronic masking? a. A post-processing function that removes nonclinical image data b. A method of increasing kVp c. A positioning technique d. A type of detector
a. A post-processing function that removes nonclinical image data
121
What is another name for electronic masking? a. Windowing b. Cropping c. Rescaling d. Filtering
b. Cropping
122
Electronic masking can be performed in what way? a. Only manually b. Only automatically c. Either manually or automatically d. Only by the radiologist
c. Either manually or automatically
123
What is the main purpose of electronic masking? a. Increase spatial resolution b. Reduce patient dose c. Eliminate extraneous brightness at image edges d. Improve positioning accuracy
c. Eliminate extraneous brightness at image edges
124
What visual issue does electronic masking help reduce? a. Motion blur b. Veil glare c. Quantum mottle d. Scatter radiation
b. Veil glare
125
What type of data is removed by electronic masking? a. Clinical anatomy b. Exposure index data c. Nonclinical image data d. Resolution data
c. Nonclinical image data
126
What effect does electronic masking have on image resolution? a. Increases resolution b. Decreases resolution c. Eliminates resolution d. No impact on resolution
d. No impact on resolution
127
Which of the following is TRUE about electronic masking? a. It replaces collimation b. It physically reduces the x-ray beam c. It is performed after image acquisition d. It reduces scatter production
c. It is performed after image acquisition
128
Is electronic masking a substitute for collimation? a. Yes, it replaces collimation b. Only in digital systems c. No, it is not a substitute d. Only when using CR
c. No, it is not a substitute
129
What can electronic masking affect in older CR/DR systems? a. Image resolution b. Exposure Index (EI) accuracy c. Patient positioning d. Beam quality
b. Exposure Index (EI) accuracy
130
Why is masking potentially problematic for radiologists? a. It increases noise b. It reduces resolution c. It can hide incidental findings d. It changes contrast
c. It can hide incidental findings
131
What are incidental findings? a. Planned anatomy in the image b. Artifacts from motion c. Exposure errors d. Unexpected findings outside the main area of interest
d. Unexpected findings outside the main area of interest
132
According to ASRT, what should NOT be done with electronic masking? a. Use it automatically b. Adjust brightness c. Remove nonclinical data d. Crop out areas that were exposed and should be interpreted
d. Crop out areas that were exposed and should be interpreted
133
What is the main ethical concern with electronic masking? a. Increased patient dose b. Loss of diagnostic information c. Reduced image contrast d. Equipment damage
b. Loss of diagnostic information
134
Why should masking be used cautiously in digital imaging? a. It affects kVp b. It reduces mAs c. It can misrepresent the original exposure field d. It increases scatter
c. It can misrepresent the original exposure field
135
What is the ASRT stance on masking clinically relevant areas? a. It is recommended b. It is optional c. It depends on EI d. It should not be done
d. It should not be done
136
What are medical radiographs considered in healthcare? a. Educational tools b. Legal documents c. Temporary images d. Research material
b. Legal documents
137
Why are radiographs considered legal documents? a. They are used for billing b. They are easily editable c. They are part of the patient’s medical record d. They are only used for teaching
c. They are part of the patient’s medical record
138
What does a medical image represent? a. Only bone structures b. A pictorial record of anatomy and medical condition c. Only positioning accuracy d. Exposure technique
b. A pictorial record of anatomy and medical condition
139
What is meant by a radiograph being a 'pictorial record'? a. It replaces written documentation b. It visually represents the patient’s condition c. It improves resolution d. It reduces exposure time
b. It visually represents the patient’s condition
140
A radiographic image represents the patient at what point in time? a. Over several days b. A projected future condition c. A single moment in the patient’s medical timeline d. An average of multiple exposures
c. A single moment in the patient’s medical timeline
141
Why is it important that radiographs reflect a single moment in time? a. It allows editing later b. It ensures consistent brightness c. It captures a specific clinical condition at that time d. It reduces noise
c. It captures a specific clinical condition at that time
142
What implication does the 'legal document' concept have for technologists? a. Images can be altered freely b. Accuracy and integrity must be maintained c. Only radiologists are responsible d. Images can be deleted if unclear
b. Accuracy and integrity must be maintained
143
Which statement best describes a radiologic image? a. A flexible visual estimate b. A temporary diagnostic guess c. A fixed legal and clinical record of a moment in time d. A modifiable teaching tool
c. A fixed legal and clinical record of a moment in time
144
What determines the accuracy of medical image interpretation? a. Only radiologist experience b. Quality of the image created c. Patient condition d. Exposure time only
b. Quality of the image created
145
What components contribute to image quality optimization? a. Only technological components b. Only human components c. Both technological and human components d. Only equipment settings
c. Both technological and human components
146
What do radiologists expect regarding department procedures? a. They are optional b. They vary per patient c. They are ignored in emergencies d. They are followed consistently
d. They are followed consistently
147
What must be done if there is a variance from standard procedures? a. Ignore it b. Delete the image c. Explain it completely d. Repeat the exam immediately
c. Explain it completely
148
What assumption do radiologists make about how images are produced? a. They are produced quickly b. They follow manufacturer settings c. They are ALARA-compliant d. They use high mAs
c. They are ALARA-compliant
149
What is the technologist’s responsibility in image creation? a. Only positioning the patient b. Ensuring optimal image quality and adherence to protocols c. Adjusting brightness only d. Recording patient history
b. Ensuring optimal image quality and adherence to protocols
150
Why is adherence to department routines important? a. It reduces scan time b. It ensures consistency and reliability of images c. It increases brightness d. It eliminates noise
b. It ensures consistency and reliability of images
151
What happens if image quality is poor? a. Radiologist compensates easily b. Interpretation accuracy decreases c. It has no effect d. It improves contrast
b. Interpretation accuracy decreases
152
What is the purpose of electronic annotation? a. To replace lead markers b. To indicate image acquisition conditions c. To reduce exposure d. To improve resolution
b. To indicate image acquisition conditions
153
Which of the following is an example of information added by electronic annotation? a. kVp and mAs b. SID c. Upright or decubitus position d. Grid ratio
c. Upright or decubitus position
154
Can electronic annotation replace lead markers for laterality (Lt/Rt)? a. Yes, always b. Only in digital systems c. Only if approved by radiologist d. No, absolutely not
d. No, absolutely not
155
What is the main advantage of electronic annotation? a. Improves contrast b. Reduces noise c. Convenient and easy to apply (cut and paste) d. Eliminates artifacts
c. Convenient and easy to apply (cut and paste)
156
Why is electronic annotation valuable to radiologists? a. It increases resolution b. It provides acquisition details c. It reduces scan time d. It replaces exposure charts
b. It provides acquisition details
157
According to ASRT best practice, how should laterality be indicated? a. Electronically after exposure b. With written notes c. Using lead markers at time of exposure d. Using PACS tools
c. Using lead markers at time of exposure
158
What is the key limitation of electronic annotation? a. It cannot be edited b. It cannot improve brightness c. It cannot indicate exposure factors d. It cannot replace anatomical side markers
d. It cannot replace anatomical side markers
159
Why must lead markers be placed during exposure? a. To improve resolution b. To reduce scatter c. To ensure permanent and legal identification of laterality d. To adjust brightness
c. To ensure permanent and legal identification of laterality
160
What are image artifacts in digital radiography? a. Extraneous information that does not represent the patient b. Improved image features c. Increased resolution areas d. Normal anatomical variations
a. Extraneous information that does not represent the patient
161
What is diagnostic yield? a. The amount of useful diagnostic information obtained b. The exposure index value c. The level of scatter radiation d. The image brightness
a. The amount of useful diagnostic information obtained
162
What does diagnostic efficacy refer to? a. Image fidelity (accuracy of representation) b. Exposure time c. Patient positioning d. SID adjustments
a. Image fidelity (accuracy of representation)
163
What does LCR stand for in image quality? a. Low-contrast resolution b. Linear contrast ratio c. Light capture rate d. Level contrast range
a. Low-contrast resolution
164
What happens to diagnostic yield between different imaging modalities? a. It remains constant b. It varies between modalities c. It only depends on kVp d. It is unrelated to imaging
b. It varies between modalities
165
Are artifacts avoidable in medical imaging? a. Yes, always b. No, they are a known fact of image production c. Only in DR systems d. Only with high kVp
b. No, they are a known fact of image production
166
Why are artifacts important to recognize? a. They improve contrast b. They may mimic or obscure pathology c. They reduce exposure time d. They increase resolution
b. They may mimic or obscure pathology
167
What is the main issue with artifacts in an image? a. They enhance diagnostic value b. They increase brightness c. They misrepresent the patient’s condition d. They reduce scan time
c. They misrepresent the patient’s condition
168
Why do digital detectors produce more gray shades? a. Wider dynamic range of photon energies b. Lower kVp settings c. Increased SID d. Reduced mAs
a. Wider dynamic range of photon energies
169
What property of digital detectors contributes to subtle gray tones? a. Spectral response b. Grid ratio c. Focal spot size d. Collimation
a. Spectral response
170
What is the result of increased gray shades in DR images? a. More subtle differences in tissue contrast b. Increased motion blur c. Reduced image brightness d. Increased artifacts only
a. More subtle differences in tissue contrast
171
Which of the following can create artifacts on DR images? a. Clothing and hair braids b. kVp settings c. SID changes d. Exposure index
a. Clothing and hair braids
172
Why are artifacts like sheets or positioning aids important to recognize? a. They may appear as pathology b. They improve contrast c. They reduce noise d. They increase resolution
a. They may appear as pathology
173
What should be done if unexpected findings appear on an image? a. Ignore them b. Repeat exposure immediately c. Explain them to the radiologist d. Delete the image
c. Explain them to the radiologist
174
What is one disadvantage of the wide dynamic range in DR systems? a. Increased chance of subtle artifacts being visible b. Reduced image detail c. Increased patient dose always d. Decreased resolution
a. Increased chance of subtle artifacts being visible
175
What is low-contrast resolution (LCR) primarily associated with? a. Ability to distinguish subtle differences in gray shades b. Ability to see edges sharply c. Patient positioning accuracy d. Exposure time
a. Ability to distinguish subtle differences in gray shades
176
What is the primary cause of data drop artifacts in DR? a. DEL saturation due to excessive exposure b. Underexposure c. Motion distortion d. Poor collimation
a. DEL saturation due to excessive exposure
177
What is the main effect of a data drop artifact? a. Increased resolution b. Misrepresentation of patient information c. Reduced noise d. Improved contrast
b. Misrepresentation of patient information
178
Which condition leads to DEL saturation? a. Low kVp b. Insufficient exposure c. Excessive exposure d. Large SID
c. Excessive exposure
179
What is one post-processing method used to correct data drop artifacts? a. Histogram analysis b. Pixel recovery c. Window leveling d. Annotation
b. Pixel recovery
180
Which of the following is a physical method to reduce data drop artifacts? a. Electronic masking b. Increasing SID c. Collimation d. Reducing grid ratio
c. Collimation
181
What is tissue bolusing used for in data drop correction? a. Increase exposure to underexposed areas b. Reduce motion c. Improve resolution d. Eliminate scatter
a. Increase exposure to underexposed areas
182
Which method should NOT be used to correct data drop artifacts? a. Masking b. Collimation c. Filtration d. Pixel recovery
a. Masking
183
What is the purpose of compensating filtration? a. Equalize exposure across the image b. Increase noise c. Reduce brightness only d. Improve positioning
a. Equalize exposure across the image
184
What is a 'ghosting' or 'phantom image' artifact? a. A residual image from a previous exposure b. Motion blur from patient movement c. Scatter radiation pattern d. Low contrast resolution
a. A residual image from a previous exposure
185
What typically causes scratches or tears artifacts? a. Damaged imaging plate or detector surface b. Incorrect kVp c. High SID d. Patient motion
a. Damaged imaging plate or detector surface
186
What do opaque spots on an image usually indicate? a. Dirt or debris on the detector b. Overexposure c. Motion artifact d. High contrast
a. Dirt or debris on the detector
187
What causes white lines on a digital image? a. Detector or reader malfunction b. Low mAs c. Increased SID d. Patient positioning error
a. Detector or reader malfunction
188
What are light spots on an image most often caused by? a. Dust or contamination on the system b. Excessive kVp c. Motion distortion d. Collimation error
a. Dust or contamination on the system
189
What does 'drop out' refer to in digital artifacts? a. Loss of signal in part of the image b. Increased brightness c. Improved resolution d. Reduced noise
a. Loss of signal in part of the image
190
What is backscatter fogging caused by? a. Radiation scattering back to the detector b. Low exposure c. High grid ratio d. Patient movement
a. Radiation scattering back to the detector
191
What are CR transport artifacts associated with? a. Issues in the plate handling system b. Incorrect SID c. High mAs d. Poor collimation
a. Issues in the plate handling system
192
What is required for all image artifacts according to best practice? a. Ignore them if minor b. Adjust brightness only c. Provide explanation and method for removal d. Repeat exposure immediately
c. Provide explanation and method for removal