the strongest reflection occurs when the beam strikes an interface at?
90 degrees
the amount of time it takes US to travel 1 centimeter into tissue and 1cm back to the transducer
13 microsecond rule
the average speed of the US through soft tissue
1540 m/sec
the difference between the wave frequency directed into the tissues and what returns.
doppler shift
the doppler angle of insonation for vascular application
60 degrees or less
-two piezoelectric crystals
-has no range or depth resolution
-does not alia
continuous wave doppler
two types of continuous wave doppler
analog
digital
uses a zero-crossing frequency detector
analog
what two ways can you solve for the doppler shift equation
-solve for frequency
-solve for velocity
what are the two techniques employed for doppler analysis of blood flow?
-continuous wave
-pulsed wave
what type of testing is continuous wave used for?
-non-imaging
-physiologic testing
continuous wave doppler
-two piezoelectric crystals
-has no range or depth resolution (signals obtained from anywhere along the line of sight are displayed)
-does not alias
counts the number of times the input signal crosses through zero (the baseline) within a given period of time, estimates and displays the frequencies present in the reflected wave
zero-crossing frequency detector
drawbacks of analog doppler
-noise
-decreased sensitivity
analyzes and displays all the frequencies moving through the sample area
fast fourier transfer
what are the pros of digital (spectral analysis)
-more accurate
-better sensitivity
-less noise
a single pulse of US is directed into the tissues, and the instrument waits for it to return before sending another pulse
pulse-listen cycle
does pulsed doppler have range and depth resolution
yes
the number of pulses per unit of time sent into the tissue
pulse repetition frequency
small footprint, triangular US beam shape, resolution decreases in far field as scan lines diverge, not able to beam steer, typical frequency 1-5 MHz
sector transducer
primary transducer for vascular application, large footprint, rectangular US beam, can beam steer, 5-10 MHz
linear transducer
large footprint, trapezoidal US beam shape, resolution decreases in far field, not able to beam steer, typical frequency 2-5 MHz
curvilinear transducer
number of cycles or vibrations per second
frequency
what is the typical diagnostic ultrasound range in frequency
2MHz to 15 MHz