Quality Control
Clinical Laboratory Test
Factors that contribute to TOTAL ERROR
sample collection, transportation, processing, and storage
Pre-analytic variability
test performance
Analytic variability
drugs or metabolic components
Interfering substances
Surrogate samples
- SD is a measure of expected imprecision in a measurement procedure when it is performing correctly
Imprecision
Systematic Bias
a method which eliminates systematic bias (within uncertainty limits)
-Correct calibration
Accuracy
Precision
Primary Purpose of Measuring QC Samples
The performance of a new method can be assessed for accuracy by _____
Assaying patient specimens or interlaboratory survey materials with known values
Assaying Patient Specimens
used to refer to an average systematic bias that may be present in a given method
Trueness
most often performed by the laboratory using calibrator materials provided by the method or instrument manufacturer
Calibration
require calibration or calibration verification at least every 6 months, or more frequently if recommended by the method manufacturer
Clinical Laboratory Improvement Act (CLIA) regulations, Section 493.1255
Statistical Quality Control
Standard Operating Procedures (SOPs)
If the QC result is within acceptable limits of the known value
the measurement procedure is verified to be stable and results for patient samples can be reported
If a QC result is not within acceptable limits
the measurement procedure is not performing correctly, results for patient samples are not reported, and corrective action is necessary
- most common presentation for evaluating QC results
Levey-Jennings (Levey, 1950)
±1 SD
68.3% of observations