There are minimal superimpositional structures as the image can be reconstructed later.
The conventional slice by slice and the volume data acquisition.
Some x-rays will be transmitted and contribute to the image as useful signals, some will be absorbed which will also be useful as useful signals by being partly absorbed and appearing brighter and some will be scattered.
The Hounsfield scale is a scale used to determine what type of body part is shown based on its shade of grey which is determined by its density. It is derived from the relative sizes of x ray attenuation of the tissue compared to water.
Window width controls the contrast while the window level controls the density.
Lung tissue
X-ray tube – Releases the x-ray photons,
gantry – holds the x-ray tube in addition to the detector array and the high voltage generator,
filters – determine which range of x-ray photon energies can pass through,
collimators – opens or closes to control scatter radiation,
detectors – arranged in an arc to detect x-rays that passed through the body,
moving couch – holds the patient in a supine position.
Slip rings allowed electricity to be conducted by brushes that touch the rings which allowed continuous detector array rotation and introduced volume scanning.
MDCT(Multi detector CT) allows simultaneous acquisition of thick and thin slices over large volumes. Thick slices produce low image noise while thin slices allow high quality multi-planar reconstructions (MPRs)
Large anode discs, high speed rotors, direct oil cooling of the anode, and advanced manufacturer design to use focal spot cooling algorithms
High dynamic range, high efficiency, robust and stable, fast response and little after glow
Collimators reduce the patient dose and image quality
Alteration Increase KVP - effect on spatial resolution decreases - effect on SNR increases - effect on radiation dose increases
Decrease mAs
Decrease slice thickness
Decrease FOV
Increase the matrix from 512x512 to 1024x1024
use of iterative reconstruction rather than filtered back projection
Increasing pitch
type of artifact is demonstrated in the images A-C and suggest a remedial solution