Capacitance
Storage of charge
Inductance
Storage of current flow
Ampere
Shortened to “amp”
-Unit of electric current (amount of electrical flow)
-Measures how much electricity is moving through the circuit
A (ampere) —> mA (milliamp) —> uA (microamp)
1 A = 1,000 mA
1 mA = 1,000 uA
1 A = 1,000,000 uA
Charge
Electrical particles
Ohm
Unit that measures electrical resistance
Ohm —> kOhm (kilo-ohm) —> MOhm (mega-ohm)
1 Ohm = 1,000 kOhms
1 kOhm = 1,000 MOhms
1 MOhm = 1,000,000 Ohms
Volt
-Electrical pressure
-The force that pushes electricity through a wire
-The electrical pressure used to stimulate the heart muscle
Hertz
Measures how often something repeats in 1 second
1 Hz = 1 cycle per second
1 Hz = 60 beats per minute
Capacitor reforming
-Electrolytic capacitors develop relatively large leakage currents over time and this can be reduced by recharging/reforming the capacitor
-Failure to reform with sufficient frequency can cause significant delays in the first shock during therapy delivery (subsequent therapies are fine)
-ICDs perform automatic capacitors reformations periodically
-It’s a charge-discharge cycle
Resistor
-Component that slows down the flow of electricity
-Limits current
-Creates resistance
-Helps control current delivered to the heart and energy used from the battery
Diode (Zener diode)
-Designed to protect the circuitry from high external voltages (like what occurs with defibrillation)
-When the input voltage presented to the pacemaker exceeds the Zener voltage, the excess voltage is shunted back through the leads to the myocardium
Impedance
-Term applied to the resistance to current flow in the pacing system
-Implies inclusion of all factors that contribute to current flow impediment
-Note that resistance is technically not interchangeable as resistance doesn’t include the effects of storage of charge or storage of current flow
Measured in ohms
Normal: 300-1200
Lead fracture (conductor failure): >2000
Loose set screw: abnormally high impedances
Insulation break: <200
Normal shock impedance: 25-100 ohms
Ohm’s law
V = I x R
Voltage = Current (amps or mA) x Resistance/Impedance (ohms)
-The lower the pacing impedance, the greater the current flow
-The higher the pacing impedance, the lower the current flow
Voltage
Pacing output strength
Measured in volts
V (voltage in volts)= I (current in amps or mA) x R (resistance/impedance in ohms)
Current
Flow of electricity delivered to the myocardium
Measured in amps or mA
I (current in amps or mA) = V (voltage in volts) / R (resistance or impedance measured in ohms)
Resistance
The opposition in the lead system
Measured in ohms
R (resistance or impedance in ohms) = V (voltage measured in volts) / I (current measured in amps or mA)
Watt
-Total energy being used
-Measurement of power
Watts = Volts x Amps
Farad
Unit used to measure capacitance
Most medical device capacitors are in microfarads (uF)
How much electricity a capacitor can hold
(If a capacitor is a bucket, the Farad is the size of the bucket)
Slew rate
-How fast voltage can change
-Represents the maximal rate of change of an electrical potential between the sensing electrodes
-Measured in volts per second
-Slew rate (V/s) = Voltage (in mV) / time (in ms)
-Should be > 0.5 V/s
-The higher the frequency content the higher the slew rate and the more likely the signal will be sensed
-The slow broad signals (T wave) have a lower slew rate and lower frequency density and are less likely to be sensed
Amplitude
-The difference in voltage recorded between 2 electrodes
-Measured in mV
Normal ranges:
Atrial: 1.5-5 mV
Ventricular: 5-25 mV
Sensing amplifier
-ICDs must amplify ventricular EGMs 10 times more than brady devices that use fixed gain sensing (because of amplitude of VF signals)
-Each sensing cycle beginning after the end of the blanking periods starts at high sensing threshold (least sensitive) and increases until new depolarization is reached or predetermined minimum threshold has been reached
-Sensitivity is minimal immediately after a depolarization and maximal late in diastole
Afterpotentials
-The excess of positive charge surrounding the cathode after termination of the pulse stimulus
-Most likely to be sensed when programmed to high voltage and long pulse duration in combination with maximal sensitivity
-Refractory and blanking periods have helped to prevent the pacemaker from reacting to afterpotentials
-Can cause crosstalk
Far field
2 definitions:
-EGMs recorded between large, widely spaced electrodes, one of which is not in contact with the heart (example: shock electrodes)
-signals from a remote source recorded on an intracardiac bipole, such as a far field R wave recorded from an atrial bipole (Signals not generated by the tissue the lead electrode is in contact with)
Sensing filters
Maximum frequency densities in sinus rhythm:
-Atrium: 80-100 Hz
-Ventricle: 10-30 Hz
Myopotential frequencies:
-10-200 Hz
High pass filter:
-Removes very low frequency signals (respiration artifact, motion artifact)
-Allows fast depolarization signals
Low pass filter:
-Removes very high frequency signals (electromagnetic interference, muscle noise)
Band pass filter:
-Allows only the frequency band typical for cardiac signals
-Filters out high and low frequency noise
Cross talk
-Atrial afterpotentials of sufficient strength and duration to be sensed by the ventricular channel
-Atrial output pulse is detected by the ventricular sensing channel
-Can result in inappropriate ventricular inhibition
-Particularly seen in unipolar systems
Ways to mitigate this:
-Adjust ventricular blanking periods
-Ventricular safety pacing algorithms
-Reduce atrial output amplitude
-Adjust ventricular sensitivity settings