Device Concepts Flashcards

(39 cards)

1
Q

Capacitance

A

Storage of charge

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Inductance

A

Storage of current flow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Ampere

A

Shortened to “amp”

-Unit of electric current (amount of electrical flow)
-Measures how much electricity is moving through the circuit

A (ampere) —> mA (milliamp) —> uA (microamp)

1 A = 1,000 mA
1 mA = 1,000 uA
1 A = 1,000,000 uA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Charge

A

Electrical particles

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Ohm

A

Unit that measures electrical resistance

Ohm —> kOhm (kilo-ohm) —> MOhm (mega-ohm)

1 Ohm = 1,000 kOhms
1 kOhm = 1,000 MOhms
1 MOhm = 1,000,000 Ohms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Volt

A

-Electrical pressure
-The force that pushes electricity through a wire
-The electrical pressure used to stimulate the heart muscle

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Hertz

A

Measures how often something repeats in 1 second

1 Hz = 1 cycle per second

1 Hz = 60 beats per minute

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Capacitor reforming

A

-Electrolytic capacitors develop relatively large leakage currents over time and this can be reduced by recharging/reforming the capacitor
-Failure to reform with sufficient frequency can cause significant delays in the first shock during therapy delivery (subsequent therapies are fine)
-ICDs perform automatic capacitors reformations periodically
-It’s a charge-discharge cycle

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Resistor

A

-Component that slows down the flow of electricity
-Limits current
-Creates resistance
-Helps control current delivered to the heart and energy used from the battery

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Diode (Zener diode)

A

-Designed to protect the circuitry from high external voltages (like what occurs with defibrillation)
-When the input voltage presented to the pacemaker exceeds the Zener voltage, the excess voltage is shunted back through the leads to the myocardium

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Impedance

A

-Term applied to the resistance to current flow in the pacing system
-Implies inclusion of all factors that contribute to current flow impediment
-Note that resistance is technically not interchangeable as resistance doesn’t include the effects of storage of charge or storage of current flow

Measured in ohms

Normal: 300-1200
Lead fracture (conductor failure): >2000
Loose set screw: abnormally high impedances
Insulation break: <200

Normal shock impedance: 25-100 ohms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Ohm’s law

A

V = I x R

Voltage = Current (amps or mA) x Resistance/Impedance (ohms)

-The lower the pacing impedance, the greater the current flow
-The higher the pacing impedance, the lower the current flow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Voltage

A

Pacing output strength

Measured in volts

V (voltage in volts)= I (current in amps or mA) x R (resistance/impedance in ohms)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Current

A

Flow of electricity delivered to the myocardium

Measured in amps or mA

I (current in amps or mA) = V (voltage in volts) / R (resistance or impedance measured in ohms)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Resistance

A

The opposition in the lead system

Measured in ohms

R (resistance or impedance in ohms) = V (voltage measured in volts) / I (current measured in amps or mA)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Watt

A

-Total energy being used
-Measurement of power

Watts = Volts x Amps

17
Q

Farad

A

Unit used to measure capacitance

Most medical device capacitors are in microfarads (uF)

How much electricity a capacitor can hold

(If a capacitor is a bucket, the Farad is the size of the bucket)

18
Q

Slew rate

A

-How fast voltage can change
-Represents the maximal rate of change of an electrical potential between the sensing electrodes

-Measured in volts per second
-Slew rate (V/s) = Voltage (in mV) / time (in ms)

-Should be > 0.5 V/s

-The higher the frequency content the higher the slew rate and the more likely the signal will be sensed
-The slow broad signals (T wave) have a lower slew rate and lower frequency density and are less likely to be sensed

19
Q

Amplitude

A

-The difference in voltage recorded between 2 electrodes
-Measured in mV

Normal ranges:
Atrial: 1.5-5 mV
Ventricular: 5-25 mV

20
Q

Sensing amplifier

A

-ICDs must amplify ventricular EGMs 10 times more than brady devices that use fixed gain sensing (because of amplitude of VF signals)
-Each sensing cycle beginning after the end of the blanking periods starts at high sensing threshold (least sensitive) and increases until new depolarization is reached or predetermined minimum threshold has been reached
-Sensitivity is minimal immediately after a depolarization and maximal late in diastole

21
Q

Afterpotentials

A

-The excess of positive charge surrounding the cathode after termination of the pulse stimulus
-Most likely to be sensed when programmed to high voltage and long pulse duration in combination with maximal sensitivity
-Refractory and blanking periods have helped to prevent the pacemaker from reacting to afterpotentials
-Can cause crosstalk

22
Q

Far field

A

2 definitions:
-EGMs recorded between large, widely spaced electrodes, one of which is not in contact with the heart (example: shock electrodes)
-signals from a remote source recorded on an intracardiac bipole, such as a far field R wave recorded from an atrial bipole (Signals not generated by the tissue the lead electrode is in contact with)

23
Q

Sensing filters

A

Maximum frequency densities in sinus rhythm:
-Atrium: 80-100 Hz
-Ventricle: 10-30 Hz

Myopotential frequencies:
-10-200 Hz

High pass filter:
-Removes very low frequency signals (respiration artifact, motion artifact)
-Allows fast depolarization signals

Low pass filter:
-Removes very high frequency signals (electromagnetic interference, muscle noise)

Band pass filter:
-Allows only the frequency band typical for cardiac signals
-Filters out high and low frequency noise

24
Q

Cross talk

A

-Atrial afterpotentials of sufficient strength and duration to be sensed by the ventricular channel
-Atrial output pulse is detected by the ventricular sensing channel
-Can result in inappropriate ventricular inhibition
-Particularly seen in unipolar systems

Ways to mitigate this:
-Adjust ventricular blanking periods
-Ventricular safety pacing algorithms
-Reduce atrial output amplitude
-Adjust ventricular sensitivity settings

25
Sensing
Measured amplitude of intrinsic cardiac electrograms as detected by the implanted device sensing circuitry Normal ranges: -Atrial 1.5-5 mV -Ventricular 5-25 mV
26
Threshold
-Definition: minimal amount of energy required to cause successful depolarization of the myocardium in contact with the electrode (and propagation of that depolarization to surrounding myocardium) -At very small pulse widths = small changes result in significant changes on the pulse amplitude -At longer pulse widths = small changes have little effect on the pulse amplitude -May vary slightly with circadian pattern, increasing during sleep and decreasing during the day (reflection of autonomic tone changes?) -May rise after eating, hyperglycemia, hypoxemia, acute viral illness, or electrolyte fluctuations -Fluctuations are typically minimal except severe hypoxemia or electrolyte abnormalities can lead to loss of capture
27
Myopotentials
Electrical signals produced by skeletal muscle activity -Examples: pectoral muscles, diaphragm, arm/shoulder muscles
28
Electromagnetic interference (EMI)
-When external electrical or magnetic signals interfere with the sensing function of a pacemaker/ICD -Can cause oversensing, pacing inhibition, inappropriate ICD shocks, noise on intracardiac electrograms -Common sources: MRI, electocautery during surgery, arc welding, high voltage power lines, radiation therapy (can cause device reset, loss of telemetry, device failure), TENS unit
29
Charge time
-Measure of both battery and capacitor function -An ICD charges it capacitor from the battery until it reaches the programmed shock voltage -Generally <10 seconds -Charge time increases when: battery ages, high shock energy is programmed, or internal resistance rises
30
ERI
Measured in the unloaded state
31
EOL
End of life -Voltage is the minimum loaded voltage -Recorded while the battery is maximally stressed Most devices lose telemetry and programming capabilities Devices frequently revert to a fixed high-output pacing mode to maintain patient safety (benefit of lithium based power cells in current devices)
32
Exit block
-When the capture threshold exceeds the programmed output of the pacemaker and loss of capture occurs -Consequence of lead maturation, results from progressive rise in thresholds over time
33
BOL
Beginning of life Cell generates approximately 2.8 V
34
Chronaxie
The threshold pulse width duration at twice the rheobase voltage (Example: If rheobase is 1.1 V then chronaxie would be whatever the pulse width duration is at 2.2 V) Typical values for modern steroid eluting leads range from 0.24-0.5 ms -It approximates the point of minimum threshold energy (microjoules) required for myocardial depolarization -Can be helpful in determine optimal pulse width -Programming the pulse width at or near chronaxie: -Achieves the most energy efficient pacing configuration -Maximizes battery longevity while maintaining reliable myocardial capture
35
Rheobase
Minimum voltage (V) amplitude required to achieve myocardial capture at an infinitely long pulse duration (ms) on the strength duration curve Represents the theoretical lowest possible capture threshold regardless of pulse width
36
NBG code
5 position standardized system to describe pacemaker programming modes Developed by the (N)orth american society of pacing and electrophysiology and (B)ritish pacing and electrophysiology (G)roup 1 - Chamber paced O (none). A (atrium). V (ventricle). D (dual). 2 - Chamber sensed O (none). A (atrium). V (ventricle). D (dual, A+V) 3 - Response to SENSED events O (none). T (triggered). I (inhibited). D (dual, T+I) 4 - Rate modulation capability O (none). P (simple programming). M (multi programmable). C (communicating) 5 - Multisite pacing capability O (none). A (atrium). V (ventricle). D (dual, A+V)
37
Signal related undersensing
Failure to sense a depolarization signal due to insufficient amplitude or frequency
38
Functional undersensing
-Failure to sense signals that have sufficient amplitude and frequency -Most commonly occurs when sensed events time in a device blanking or refractory period
39
Near field
Recorded from closely spaced bipoles in contact with the source myocardium