What is the mode most likely being used in the ECG below:
a. AAI
b. VOO
c. DDD
d. VVI
d. VVI
What is the underlying rhythm in the ECG?
a. Atrial flutter
b. Complete heart block
c. NSR
d. Atrial fibrillation
b. Complete heart block
What is the correct interpretation?
a. Normal sinus with atrial pacing
b. Normal sinus with ventricular pacing with failure to capture
c. Normal sinus with atrial pacing with failure to capture
d. Normal sinus with ventricular pacing
b. Normal sinus with ventricular pacing with failure to capture
What is the correct interpretation?
a. Atrial sensed-ventricular sensed
b. Atrial sensed-ventricular paced
c. Atrial paced-ventricular paced
d. Atrial paced-ventricular sensed
c. Atrial paced-ventricular paced
What is the most likely explanation for the narrower paced complexes in the following ECG?
a. CRT device
b. Not being paced at all
c. Only atrial pacing occurring
d. ICD device
a. CRT device
TRUE OR FALSE
CRT devices are programmed to pace as close to 100% as possible.
TRUE
What is the correct interpretation?
a. Atrial paced-ventricular sensed
b. Atrial sensed-ventricular paced
c. Atrial paced-ventricular paced
d. Atrial sensed-ventricular sensed
c. Atrial paced-ventricular paced
What is the correct interpretation?
a. Atrial fibrillation with AS-VP with AP and PACs
b. Normal sinus with AS-VP and failure to capture
c. Normal sinus with AS-VP with occasional AP and PACs
d. Atrial fibrillation with AS-VP with AP and PACs
c. Normal sinus with AS-VP with occasional AP and PACs
What is the correct interpretation?
a. Normal sinus rhythm with ventricular paced beats and pseudofusion beats
b. Normal sinus rhythm with ventricular paced beats
c. Atrial fibrillation with ventricular paced beats and pseudofusion beats
d. Atrial fibrillation with ventricular paced beats
c. Atrial fibrillation with ventricular paced beats and pseudofusion beats
What is the correct interpretation?
a. Atrial pacing with a run of PMT
b. Dual chamber pacing
c. Ventricular pacing
d. Dual chamber pacing with a run of PMT
d. Dual chamber pacing with a run of PMT
If the indication is 24 hours post-implant for the above ECG, what is the most likely reason for no tracing in V6?
a. Lead wire fell off
b. Technologist error
c. No access to the correct position
d. Damaged lead wire
c. No access to the correct position
TRUE OR FALSE
PMT is a re-entry tachycardia.
True
What is the correct interpretation?
a. Atrial fibrillation with ventricular pacing
b. Normal sinus with atrial pacing
c. Atrial fibrillation with atrial pacing
d. Normal sinus with ventricular pacing
b. Normal sinus with atrial pacing
What is the correct interpretation?
a. Normal sinus with ventricular pacing
b. Normal sinus with atrial pacing
c. Normal sinus with atrial pacing and PACs
d. Normal sinus with ventricular pacing with PACs
c. Normal sinus with atrial pacing and PACs
What is the underlying rhythm of the following ECG?
a. Atrial flutter
b. Normal sinus rhythm with left bundle branch block
c. Atrial fibrillation
d. Normal sinus rhythm
c. Atrial fibrillation
TRUE OR FALSE
Infection will usually present itself within a week of implant.
False
TRUE OR FALSE
ICM are implanted for arrhythmias that have not been able to be documented through other cardiac monitors.
True
What is the most appropriate mode for a patient with complete heart block?
a. AAI
b. VVI
c. DDD
d. VVIR
c. DDD
What is most appropriate mode for a patient with atrial fibrillation and chronotropic incompetence?
a. VVI
b. DDD
c. VVIR
d. AAI
c. VVIR
Failure to capture refers to which of the following?
a. The device does not sense the appropriate cardiac impulses
b. The device senses non-cardiac impulses
c. A device doesn’t know when to turn off
d. The generator is unable to deliver sufficient energy to cause depolarization
d. The generator is unable to deliver sufficient energy to cause depolarization
Temporary pacers may be used in the following situations, except:
a. Lyme disease
b. Drug overdose
c. Ventricular fibrillation
d. Complete heart block
c. Ventricular fibrillation
Threshold testing is done to make sure there is an adequate safety margin to ensure capture. Another reason is:
a. to check impedance
b. So the patient will feel like they came to the clinic for a reason
c. So that the sensitivity can be adjusted
d. To keep the energy output as low as possible to conserve battery
d. To keep the energy output as low as possible to conserve battery
TRUE OR FALSE
The downside of VDD pacing is that it only senses the ventricle.
False
MATCH
The maximum rate that the device will track
a) PVARP
b) UTR
c) ATP
d) PMT
e) Pacing Threshold
f) USR
g) AMS
h) Pulse width
b) UTR