QC for CR + DR
DAILY
CR-check load an unload functions of IR
CR- erase all cassettes
DR- charge batteries for wireless
QC all printers
Verify all interfaces
QC for CR + DR
WEEKLY
CR- clean and rid of debris
CR- IR clean air intakes
DR- clean all detectors
Clean monitor screens
QC for CR + DR
MONTHLY
CR- inspects for cracks/scratches
DR/CR- evaluate monitors with test pattern
Repeat reject analysis
QC for CR + DR
SEMI-ANNUAL
resolution test
Spatial accuracy- wire mesh
Uniformity- homogenoeus phantom
Erasure
Sensors
QC for FLUORO
all semi-annual
What is TG 18 QC used for?
monitor AAPM QC