what is the difference between the production of xrays in flat panel and image intensifier fluroscopu
flat panel = receives xrays from high spec medium frequency generator
image intensifier = receives xrays from regular xray tube
what is the power range of the frequency generator in flat panel fluroscopy
65-126kw
what is the exposure range of the xrays the frequency generator makes for static images and fluroscopy
static imaging = 40-150kV, 32-63mA
Fluroscopy = 40-125kV, 0.5-4mA
both have 5ms-5 second times
what is the range of pulsed images per second the can be produced by the frequency generator
1 - 8 per second
(cardiac systems up to 30/sec)
describe the process of how a flat panel fluoroscopic system works
(no electron acceleration or minification to reduce noise)
why is the caesium iodide layer thicker in flat panel than image intensifier system
for greater xray absorption/sensitivity
flat panel does not have electron acceleration to minimise noise production, so what does it do instead to reduce noise
pixel binning
what is a negative of pixel binning
reduced resolution
how can pixel binning affect frame rate
increase/faster frame rates as it reduces the amount of data to be handled
what are the 2 ways to magnify an image using the flat panel system
both fluroscopy and recorded images are dynamic methods
achieved nu pulsing xrays at several frames per second
what controls the xray pulses in fluroscopy
gridded tube, it suspends electrons ready to expose
compare the dosage of pulsed and continuous fluroscopy
pulsed fluroscopy significantly reduces dose to patient
pulsed fluroscopy does reduce dosage to patient however there are disadvantages associated, what are they
for very slow pulse rates as in pulsed fluroscopy, when several frames have been averaged to smooth out the noise what can happen to the contrast of the imageq
what is the typical fluroscopy pulse rate
1-15 frames per second
what are the 5 things you can adjust to control flat detector image quality
what are the 4 types of noise reduction filters/systems
what type of artefact/effect will you find in image intensifier but not flat panel system
no pin-cushion effect due to curvature of front of image intensifier causing blur at the edges
why is there no s-distortion found in flat panel but is in image intensifior
what is done to correct s-distortion in image intensifier system
s-distortion corrections are made to the system by the service engineer
what is another advantage that flat panel has over image intensifier system
no vignetting, where the periphery of the image is less bright than the center and has lower resolution
know that overall the flat panel has 3 major benefits over image intensifier in terms of artefacts/errors that can occur in image production
what are edge detection filters used for
edge enhancement