Changes in what lead(s) would indicate inferior wall ischemia (right coronary artery)?
II, III, aVF
Changes in what lead(s) would indicate lateral wall ischemia (circumflex branch of left coronary artery)?
I, aVL, V5-V6
Changes in what lead(s) would indicate anterior wall ischemia (left coronary artery)?
V3-V4
Changes in what lead(s) would indicate septal wall ischemia (left descending coronary artery)?
V1-V2
ST-segment changes to leads II, III, and aVF could indicate ischemia to what area of the heart?
Inferior wall (right coronary artery)
ST-segment changes to leads I, aVL, V5-V6 could indicate ischemia to what area of the heart?
Lateral wall (circumflex branch of left coronary artery)
ST-segment changes to leads V3-V4 could indicate ischemia to what area of the heart?
Anterior wall (left coronary artery)
ST-segment changes to leads V1-V2 could indicate ischemia to what area of the heart?
Septal wall (left descending coronary artery)
What two leads are the standard of monitoring for HR/arrhythmia detection and for ischemia?
Lead II for HR and arrhythmia detection. Lead V5 for ischemia.
What are the principle indicators for ischemia detection on ECG?
ST-segment elevation >/= 1mm
ST-segment depression >/= 1mm
T wave flattening or inversion
Peaked T waves
Development of Q waves
Arrhythmias
What should you do if you suspect ischemia and why?
Get a TEE so that you can look at wall motion abnormalities. Then work on your supply/demand (decrease HR, increase BP).
Changes in SBP correlate with changes in…
…myocardial O2 requirements.
Changes in DBP reflect…
…coronary perfusion pressure.
What should a well-fitted NIBP cuff bladder’s width extend to (in relation to patient’s arm)?
Bladder width should be approximately 40% of the circumference of the extremity
What should a well-fitted NIBP cuff bladder’s length extend to (in relation to patient’s arm)?
Bladder length should be sufficient to encircle at least 80% of the extremity
Potential reasons for falsely high NIBP measurements
Cuff too small
Cuff too loose
Extremity below level of heart
Arterial stiffness (HTN, PVD)
Potential reasons for falsely low NIBP measurements
Cuff too large
Extremity above level of heart
Poor tissue perfusion
Too quick deflation
What patient populations are more vulnerable to NIBP measurement complications?
Peripheral neuropathies
Arterial/Venous insufficiencies
Severe coagulopathies
Recent use of thrombolytic therapy
What are complications of noninvasive blood pressure (NIBP) measurement?
Compartment syndrome
Limb edema
Pain
Peripheral neuropathy
Petechiae and ecchymoses
Venous stasis and thrombophlebitis
List the indications for arterial line cannulation.
How is the morphology of the arterial wave form affected with different arterial catheter sites?
As the pressure wave travels from the central aorta to the periphery:
How do pressures compare between the central aorta and peripheral arterial waveforms?
Peripheral arterial waveforms have:
Potential causes of overdamped arterial pressure waveforms
Potential causes of underdamped arterial pressure waveforms