differences between 3DCRT and IMRT dosimetry- challenges for IMRT
A
IMRT has to focus on cumulated dose not on QA of a single segment
IMRT: fluence is varied during treatment so requires integrative dosimetric techniques
IMRT- sometimes independent dose calculation methods are used as QA
have to consider oblique-incidence response of clyndrical ion chambers for non-coplanar beams
Volume averaging is an issue for detectors in small fields- dose will be underestimated due to averaging over the entire size of the detector
if measuring profiles of dose distributions with rapidly spatially varying doses, the shape of the dose distribution will be blurred by volume averaging – use higher spatial resolution detector if accurate measurement of profile shape is needed
small ion chamber is more sensitive to position errors