CONTROL PANEL:
Control panel:
Equipment Variability:
Control Panel Features:
ON/OFF SWITCH
EXPOSURE TIMER
Types of Exposure Timers:
MECHANICAL TIMER
MECHANICAL TIMER
Advantages of Shorter Times:
SYNCHRONOUS TIMERS
ELECTRONIC TIMERS
Advantages:
ELECTRONIC TIMERS
EVALUATION OF TIMERS
Importance of Exposure Time:
Evaluation of Timers:
SPINNING TOP DEVICE
EXPOSURE TEST PROCEDURE:
* Select exposure factors: 1/__, 1/__, or 1/__ second.
* Image produces dashes (or dots).
* Number of dashes relates to the exposure time:
* 1/&_ second: _ dashes
* 1/30 second: _ dashes
* 1/__ second: __ dashes
* For single-phase, full-wave rectified units, one dot
per 1/___ second of exposure.
* Calculate expected dots: multiply exposure time
(seconds) by 120.
EXPOSURE TEST PROCEDURE:
* Select exposure factors: 1/60, 1/30, or 1/10 second.
* Image produces dashes (or dots).
* Number of dashes relates to the exposure time:
* 1/60 second: 2 dashes
* 1/30 second: 4 dashes
* 1/10 second: 12 dashes
* For single-phase, full-wave rectified units, one dot
per 1/120 second of exposure.
* Calculate expected dots: multiply exposure time
(seconds) by 120.
Oscilloscope
Used for both single-phase
and three-phase
generators.
* Placed in the primary x-ray
beam.
* Counts and records pulses
for evaluation.
Pulse counter device
Milliampere Selector
mA SELECTIONS:
* Radiographic units offer mA selections ranging from __ to
____ mA.
Focal Spot Size Determination:
* In many general radiographic units, the selection of mA
determines the focal spot size.
* “L” (Large) indicates selections using the ______ focal spot.
* “S” (Small) indicates selections using the _____ focal spot
mA SELECTIONS:
* Radiographic units offer mA selections ranging from 25 to
2000 mA.
Focal Spot Size Determination:
* In many general radiographic units, the selection of mA
determines the focal spot size.
* “L” (Large) indicates selections using the large focal spot.
* “S” (Small) indicates selections using the small focal spot
mA meter/mAs meter