What are the 9 components of a Machine Check
How do you check emergency equipment?
How do you check the high pressure system
1) turn off the pipeline source (H tank or DISS)
2) remove the O2 sensor and cap the inlet
3) open the E-cylinder and verify psi >1000psi
4) close E-cylinder
5) depress O2 flush
6) watch bourdon gauge for e-cylinder drop to zero
7) turn on anesthesia machine
8) listen for O2 low pressure alarm
9) reconnect pipeline or H-tank
10) ensure bourdon gauge for pipeline reads 50psi
How do you check a low pressure system?
1) test flowmeters by turning on and check for smooth operation of floats
2) attempt to create a hypoxemic mixture
3) ensure flow control valves closed and vaporizers off
4) ensure vaporizer mounted, (fill vaporizers if need), and interlocked
5) verify master switch and flow control knobs are off
6) attach suction bulb to CGO
7) collapse and squeeze bulb
8) ensure stays collapsed for 10 seconds
9) open one vaporizer at a time and repeat with each vaporizer
10) remove suction bulb
11) reconnect fresh gas hose
How to check the oxygen sensor?
How do you check the scavenging system?
How do you check the Breathing circuit
How do you check the Ventilator?
-