Secondary access interlocks
Functional
Emergency off switches
Functional
Movement interlocks (full check)
Functional
Backup timer interlock (where set by user)
Functional
Pointers (if used)
2 mm
Gantry and collimator rotation scales
+/- 0.5 degrees
Optical field size variation for different field sizes
2 mm
Isocentre quick check
2 mm diameter
Distance indication at different SSDs
2 mm
Couch movement calibration
2 mm relative
Couch vertical movement
2 mm
6DoF couch
2 mm/ 1 degree
Output calibration check
+/- 2%
Flatness and symmetry
+/- 2%
Energy check using dose ratio at gantry 0 degrees
Ratio +/- 1%