Gantry
Slip rings permit the gantry frame to rotate _____, making _____ scan modes possible
Continuously, helical
Small focal spots in CT tubes produce _____ images (better spatial resolution)
Sharper images
-because they concentrate heat onto a smaller portion of the anode they cannot tolerate as much heat
High ___ is used to increase the intensity of the beam
Kv
Filtering the x-ray beam helps to
- improves image quality (by reducing image artifacts that result from beam hardening)
Collimators
reducing scatter improves _____ and decreases patient dose
contrast resolution
Source collimator aka prepatient collimation
Pre detector collimation aka postpatient collimation
Scan field of view
-determines the size of the fan beam, which in turn, determines the number of detector elements that collect data
Capture efficiency
Ability with which the detector obtains photons that have passed through the patient
Absorption efficiency
The number of photons absorbed by the detector and is dependent on the physical properties of the detector face (ex thickness, material)
Response time
The time required for the signal from the detector to return to zero after stimulation of the detector by x-radiation so that it is ready to detect another x-ray event
Dynamic range
The ratio of the maximum signal measured to the minimum signal the detectors can measure
What is the most common material a detector is made out of?
Solid-state crystal variety
Solid state crystal detector aka scintillation detectors
Detector spacing
A small detector is important for
- scatter rejection
Characteristics of solid state crystal detectors
Characteristics of pressurized xenon gas detectors (not used in newer models for CT)
Stability
Refers to the steadiness of the detector response time
-if the system is not stable, frequent calibrations are required to render the signals useful
Response time
Refers to the speed with which the detector can detect an x-ray event and recover to detect another event
-response times should be very short
Dynamic range
“Ratio of the largest signal to be measured to the precision of the smallest signal to be discriminated”
Afterglow
The persistence of the image even after the radiation has been turned off
-CT detectors should have very low afterglow values