What is Exposure Index and range for uni x-ray tube?
Range of interest indicated by a green square surrounding the image
* Linear scale - proportional to exposure
* Normal EI range usually 250-350 (may vary depending on manufacturer)
Possible reasons for changes in Exposure Index?
Exposure field recognition error – doesn’t recognise area you have
exposed
* Unexpected material in field
* Collimation margins not detected
* Scatter
If EI is out of range:
what should you do
Look where green square is
* Look for distortion
* Look what DAP says
- If DAP is within range but EI is not, likely to be a processing error.
- If DAP not within range, could be an equipment fault. Note down
exposure factors used, DAP and EI, report to senior radiographer or RPS,
perform QA, take room out of action until issue resolved.
Radiographic Brightness
High density = low brightness, low density = high brightness.
Are you able to see trabecular patterns within the bone?
* Need to say how you would rectify it if the image is too bright or not
bright enough (e.g. windowing, repeats)
What affects Exposure
How do you manage absorbed dose?
To maintain image quality, but managed absorbed dose, the following Rule of
10 should be used:
For every 10kVp increased, the mAs should be halved – gives the same overall
exposure and image quality, EI should stay the same, but DAP may change
How does AEC work
Impact of changes with AECs
Grids
Describing “normal”
Just stating ‘no bony injury’ without further detail is not enough for this exam
(possibly is in practice). You need to base your answer on the clinical history.
Some examples of descriptions of “normal”:
* No bony injury. The glenohumeral joint is congruent. The
acromioclavicular joint remains undisrupted.
* No bony injury. Normal bone, soft tissue and joint appearances.
* Normal appearances of heart, mediastinum and lungs. No evidence of
pneumothorax. normal bony anatomy (clinical history: stabbed in right
side of chest)
* Normal appearances of bone and soft tissue. No breaks in bony cortex or
abnormal lucent/opaque areas. Joints aligned normally
Adapted technique scenarios- Wider collimation than normal
say what the normal collimation should
be, say where the obvious deformity appears to be, increasing the AOI will
ensure that the fracture is fully demonstrated on the first image, hopefully
preventing need for repeats
Adapted technique scenarios- Addition (or removal) of a grid
larger volume of tissue being imaged,
increased amount of scatter produced, will adversely affect the resultant
image if a grid is not used
Adapted technique scenarios- angling of the tube
ensure perpendicular relationship between affected
body part and IR
Adapted technique scenarios- Larger SID
if IR cannot be placed in contact with the body part being
imaged, need to increase SID to account for increased magnification.
Standard SID for CXR is 180cm because the heart naturally has space
between it and the IR so need to reduce magnification.
Adapted technique scenarios- Placing affected body part on radiolucent pad
raises affected body part
enabling posterior/anterior aspect to be visualised on the IR
Explaining differences in DAP
if exposure factors used are the same then the only thing changed is collimation.
if one projection has a smaller field of view (e.g AP lumbar spine compared to AP pelvis), reduced are exposed so reduces DAP and therefore reduced DRL will be lower.
DAP measures the beam leaving the tube, not the patient dose
DAP also does not take SID into account - remains the same.
What is exposure index and what are possible reasons for change in EI
-Range of interest indicated by a green square surrounding the image
* Linear scale - proportional to exposure
* Normal EI range usually 250-350 (may vary depending on manufacturer)
if EI is out of range, what do you do?
pregnancy pelvis x-ray exposure factors
high kV technique is needed to reduce dose to foetus
increase kV and reduce mAs to get kV over 100
overall image quality and brightness maintained even though contrast is affected but still ALARP
What do you do if there is half the volume of tissue?
half the volume of tissue requires half the exposure (e.g child’s ankle compared to adults ankle)
To halve an exposure - minus 10 kV and 1/2 the mAs
(very low kV gives very good contrast)
What do you do if there is double the volume of tissue?
double the volume of tissue requires double the exposure
add 10KV and double the mAs
For x-raying COPD how will you change the exposure from the standard exposure 85kV 2.5mAs
Lungs will be very low density, reduce the exposure because of effects of cumulative dose 75kV 1.5mAs
For x-raying pleural effusion how will you change the exposure from the standard exposure 85kV 2.5mAs
same exposure, will allow comparison with previous imaging without increasing patient dose. 85kV 2.5mAs
For x-raying surgical emphysema- how will you change the exposure from the standard exposure 85kV 2.5mAs
For x-raying surgical emphysema- reduce kV due to air in soft tissues, requires less penetrating. Keep mAs same, same number of photons to maintain image quality. 75kV 2.5mAs