pacemakers
deliver electrical stimulus to cardiac tissue triggering depolarization and contraction
categories of pacemakers
permanent pacemakers
leadless pacemaker
implanted directly into heart with no separate incision under skin
temporary pacemakers
type of temporary pacemakers
how can pacemakers pace the heart?
indications for pacemaker therapy
to manage select cardiac dysrhythmias and to enhance cardiac output
indications for pacing
what does the pacemaker system comprise of?
pulse generator (power source) and one or more pacing leads; with permanent, both are internal; when external pacemakers, a connecting cable is needed
what does the pulse generator do?
the electrical impulse generated by the ____ is transmitted via ______ which are in direct contact with ______
pulse generator; the pacing leads; either the atrial or ventricular chambers (single chamber) or both (dual chamber)
can dual chamber pulse generator be used for both single or dual chamber pacing?
yes
components of pulse generators
sense light versus pace light
sense light is normally orange and lights up when pacemaker senses an intrinsic depolarization
pace light is normally green and is activated when the pacemaker emits an impulse to pace
pacemaker settings
-HR: set according to pt’s needs to maintain adequate CO; usually 60-80bpm; measured as pulses/min to distinguish between intrinsic HR and artificially generated ones
output or mA
pacing threshold
Once a paced rhythm is achieved, the output level is noted
what is mA usually set at?
~2 to 3 times the pacing threshold, allowing for changes in the pacing threshold
sensitivity
the amount of electrical activity that needs to be generated during a patient’s intrinsic depolarization for the pacemaker to sense it (measured in mV)
why is sensitivity an important setting?
what if the mV setting is higher?
the larger the amount of electrical activity the heart has to generate during its intrinsic depolarization for the pacemaker to recognize it
high mV settings indicate?
pacemaker is less sensitive to pt’s intrinsic activity
low mV settings indicate?
pacemaker is more sensitive to pt’s activity