Anesthesia machine functions
First anesthesia performed:
Oxygen E-cylinder volume and pressure
Nitrogen E-cylinder volume and pressure
1590 liters
745 psig
Air E-cylinder volume and pressure
The O2 knob is larger & textured differently than the other gases as an added safety measure as well. True or false?
True
FiO2
O2 (L/min) + Air (L/min) (0.21)
Total Gas Flow (L/min)
Only 1 vaporizer can be used at a time for safety. There is a locking mechanism, called a ______ ______that disengages the other vaporizers once 1 is selected.
vaporizer interlock
MAC of des, sevo, iso respectively
6.0, 2.0, 1.2
Vaporizers have a ____ ____ system or a ___ ___ system.
funnel fill
key fill
Circle System Components
Basic Ventilation Modes (3)
When to change absorbent
In VCV at 500mL, what happens to the peak pressure & plateau pressure when you increase the respiratory rate from 10 to 20 while leaving tidal volume constant & I:E 1:2?
Increase both peak pressure and plateau pressure
What happens to the I:E ratio when you increase respiratory rate?
decreases
What happens to the I:E ratio when you increase tidal volume?
In volume control mode, waveform is _____ shaped.
triangular
In pressure control ventilation, waveform is ______ shaped.
rectangular
Who enforces medical gas purity?
FDA
As far as regulation goes, who does pretty much everything other than testing for gas purity?
DOT
The cylinder gases go through:
_________ was designed as a safety feature to help ensure that cylinders are not installed on the wrong cylinder yoke.
Pin Index Safety System
Only used on cylinder sizes A-E
Pressure regulators are found ____________.
directly behind the yoke
If the pressure in the gas supply module is too high, above ________psig, the pressure relief valve opens and vents the gas to atmosphere. .
approximately 75