M and M of Gas Delivery Equipment (closed claims project)
Most common outcomes = death (47%) and brain damage (29%) related to gas delivery equipment
Closed Claims related to “less than appropriate care”
includes inadequate checking of anesthesia equipment and can lead to HIGH monetary malpractice awards
Prevention and Early Detection of Machine Issues
AANA Core Values
– Patient safety
– Care for the whole patient, from a nursing perspective
– Professional excellence and personal well-being
– Healthcare policy and collaboration
– Integrity and quality in all professional and clinical settings
AANA Motto
safe and effective anesthesia care
Standard VIII
6th Vital Sign
Pre-Anesthesia Machine Check History
– 24 steps
• Research showed that these guidelines were used inconsistently and the success rates in finding faults in the anesthesia machine were poor.
1993: FDA reduced steps to 14, but still were not readily used and M and M continued to be reported due to machine malfunction until 2008
2008
UIASC Checklist Philosophy (From APSF Guidelines)
• To avoid patient harm through execution of procedures designed to assure the proper function and presence of critical anesthesia requirements including:
– Reliable delivery of oxygen at any appropriate concentration up to 100%.
– Reliable means of positive pressure ventilation.
– Backup ventilation equipment available and functioning.
– Controlled release of positive pressure in the breathing circuit.
– Anesthesia vapor delivery (if intended as part of the anesthetic plan).
– Adequate suction.
– Means to conform to standards for patient monitoring.
– Means to perform adult and pediatric Advanced Cardiac Life Support (ACLS) including the presence of emergency drugs and immediate availability of defibrillation.
2008 APSF Guidelines and Design Checklist
• Delineated what tasks could be done by anesthesia technician and those that had to be done by the anesthesia provider and those that could be double checked.
– YOU ARE ULTIMATELY RESPONSIBLE!!!
Tasks to be Completed Daily, or After a Machine is Moved or Vaporizers are Changed (15 steps)

To Be Completed Prior to Each Procedure (Subset of items in the daily checklist to be completed between cases)