SVC: ERV have to vary less than ___ for __ breaths
100mls, 3 breaths
ATS FVL Acceptability Criteria
1) Good Start: get to peak w/in 120ms or don’t blow out more than 100mls or 5% of FVR prior to peak
2) End of Test: >/= to 6sec duration (adult), 3sec (children); plateau <0.025L for >/= to 1sec; expiratory time >/= 15sec
3) No Artifact: cough w/in 1sec, glottis closure, early termination, cut-off, leaks, obstructions, mouthpiece problems
ATS FVL Repeatability Criteria
Utility of Inspiratory Curve FIF50
Info re. upper airway obstruction
Flow Impairment:
FEV1/FVC </= 70% (adult) 80% (peds)
FEV1
- normal >/= 70%
- mild obstruction >70%
- moderate obstruction 60-69%
- moderately severe 50-59%
- severe obstruction 35-49%
- very severe <35%
Volume Impairment
Evaluate FVC against the % predicted normal LLN (80-120); can’t diagnose restriction w/ spirometry alone (need TLC to confirm)
MVV Acceptability Criteria
how can you measure … volumes
dynamic: spirometer; respirometer
static: displacement
how can you measure … airflows
peak: peak flow meter
mid breath: computer and/or paper based measure
how can you measure … pressures
mouth: manometer
alveolar: manometer with NO airflow and/or esophageal manometer
LLN Definition
5% of the population with healthy lungs have lower values
LLN is ~80% of predicted for FEV1 and FVC; ~90% of predicted for FEV1/FVC ratio; ~60% predicted for FEF25-75%
VOPS
Ventilate
Obstruction
Protection
Secretions
Normal Values: MIPS & MEPS
MIP
- women: -50cmh2o
- men: -75cmh2o
MEP
- women: >80cmh2o
- men: >100cmh2o
Daily Calibration Spirometry for Accuracy
Volume: w/in 3% or +/- 0.05L from 0.5-8L
Flow: w/in 5% or 0.2 L/sec up to 14 sec
which equipment can measure flow in both directions
which equipment can only measure unidirectional flow
which equipment is neg affected by turbulent flow
which equipment works with laminar or turbulent flow
which equipment has issues with high/low flow
which equipment is affected by gas composition / temperature / humidity
which equipment is not affected by gas composition / temperature / humidity
which equipment meets ATS criteria
which equipment does not meet ATS criteria