How does the vaporizer minimize the effect of cooling on vapor pressure and vaporizer output?
temperature compensating valve adjusts ratio of vaporizing chamber flow to bypass flow and guarantees a constant vaporizer output over a wide range of temperatures
How does the vaporizer minimize the effect of cooling on vapor pressure and vaporizer output?
temperature compensating valve adjusts ratio of vaporizing chamber flow to bypass flow and guarantees a constant vaporizer output over a wide range of temperatures
why will a febrile 70 kg adult become hypoxic with an O2 flow at 250 mL/min in a closed circuit system?
VO2 for avg adult is 250 mL/min
conditions like sepsis, fever, thyrotoxicosis increase VO2
2 immediate actions if O2 pipeline supply fails
What 2 functions does drive gas on a pneumatic ventilator serve?
position of ventilator spill valve during inspiration
closed
ensures that Vt goes to the patient and not to scavenger
minimum expiratory pressure needed to open ventilator spill valve
3 cm H2O
2 complications of overuse of O2 flush valve
barotrauma
awareness
where does the high pressure system begin and end
begins at the cylinder, ends at cylinder regulators
where does the intermediate pressure system begin and end
begins at pipeline, ends at flowmeter valves
components of high pressure system
components of intermediate pressure system
where does low pressure system begin and end
begins at flowmeter tubes, ends at common gas outlet
components of low pressure system
what does the low-pressure leak test assess
integrity of the low-pressure circuit from flowmeter valves to common gas outlet
how is low pressure leak test performed
by attaching a bulb to the CGO and creating negative pressure (-65 cm H2O)
If bulb reinflates within 10 seconds = fail, leak in system
when is the low pressure leak test performed
before 1st case of day
how to perform high-pressure leak test
Close APL, pressurizing circuit to 30 cm H2O
change in pressure reading = fail, leak
what does a high pressure leak test check in a machine with vs. without a check valve
where is a check valve usually located
usually downstream from vaporizer, upstream from O2 flush
when is high pressure leak test performed
before the first case of the day
what is the SPDD model
Nomenclature describes location of each gas handling component
Supply, Processing, Delivery, Disposal
What is S in SPDD model & where is it located
what is P in SPDD model and where is it located