How do filters affect the HVL?
Increase HVL (due to increased energy of electrons)
What does iterative reconstruction do?
Alloweds for increased noise which decreases patient dose
What determines minimal thickness of a slice?
Minimal detector aperture width
How do you determine how many shades of gray are in a pixel?
2^bit
What is Pitch?
Distance Table Moved/Beam Width
What does Pitch <1 Mean?
What is the relationship between X Ray attentuation and Houndsfield units?
Increase in HU by 10 attenuates X rays by 1%
How does kEV affect HU?
Closer you are to the k edge, more PE, which means more attenuation and higher HU
How does Filtration affect HU?
Decreases HU: Increase filtration= increased average energy of the beam= less PE effect (less aborption, more transmitted)
What is Axial Acquisition?
Table is stationary and the Tube Steps and Shoots
What is helical acquisition?
Tube is constantly on and the Table Moves
-Less stair step artifact
Sharp versus smooth kernal?
Sharp: More spatial resolution and more noise
Smooth: Less spatial resolution and less noise
What is cardiac imaging best acquired?
Diastole
Prospective versus Retrospective
Prospective: Always Axial, less dose, no functional imaging
Retrospective: More dose, functional imaging
Contraindications to giving B Blocker for Coronary CTA
Contraindications to giving Nitoglycerin
How does X Ray Flux effect SNR?
-Need to increase Flux 4 X to double the SNR
How do you increase SNR?
How does Pixel Size affect Spatial and Contrast Resolution?
Small Pixel
How does slice thickness affect spatial and contrast resolution?
Increased slice thickness
What is CT Dose index (CTDI)?
Radiation dose Normalized to Beam Width
What is Weighted CTDI?
2/3 Peripheral CTDI + 1/3 Central CTDI
What is Volume CTDI?
Weight CTDI/Pitch
What is Dose Length Product?
Volume CTDI X Length of Scan in cm