How does Fluoro compare to general radiographs?
How does a spot image compare to a last image hold?
When is the source above the table and what are the side effects?
- Increase dose to the lens
How does the image intensifier work?
Why is the image inverted and Reversed on the output phosphor?
Due to point inversion (all pass through the same focusing point)
What is Brightness Gain?
-Ability of the II to convert xray signal to light and increase brightness
=Flux Gain X Minification Gain
What is Flux Gain?
What is Minification gain?
What is Conversion gain?
Efficiency of the II at making x rays into light at output surface
What happens to brightness and conversion gain with older II?
They are less efficient and require a higher dose to create the same level of brightness
When do you change image intensifier?
When the conversion gain falls below 50%
What is electronic magnification?
Making the input field of view smaller (increases dose) by magnifying the II
How does Electronic Magnification affect spatial resolution?
Increased Electronic Magnification increases spatial resolution
How do you determine if geometric magnification will increase or decrease spatial resolution?
R/F Ratio: Intrinsic Receptor Resolution/Focal Spot
If < 0.5, then it decreases resolution
If >0.5, then it increases resolution
What does double the distance from the patient to source do to dose?
Decreases its by 4 times due to inverse square law
What does collimation do to dose and scatter?
Decreases dose and scatter
What is the Automatic Exposure Rate control limit?
10 R/Min or 87 mGy/min
What is KERMA?
Kinetic energy released per unit mass
=Total Energy Deposited (Heat/Ionization)/Mass
What is Air KERMA?
- Total number of photons in the air right before you hit the skin
What is Kerma Area Product?
-Dose X Cross sectional area of X ray beam
How does electronic magnification affect KAP and Air Kerma?
- Keeps KAP the same, smaller CSA to balance increased dose
What is the Air Kerma Limit?
10 Roetngens/min or 87 mGy/min
What is high level control for air kerma limit?
20 Roetngens/Min or 176 mGy/min
What is pincushion artifact?
Bent lines in the periphery due to Large FOV