Properties of Electrons
• Have a finite range
• Rapid dose falloff
• Do not deposit dose at depth, reduces normal tissue dose
• Provide a high surface dose compared to photons
• Have a wide penumbra which increases at depth
• A range of electron energies to choose from, 6,9,12,15,18 depending on the
energies commissioned
Why use Electrons?
• Provide a uniform dose from the surface to approx 6cm depending on electron
energy
• Useful in treating skin, nose, ears, chestwall, eyelids, scalp, limbs whilst sparing
normal tissue
Percentage Depth Dose
• The shape of the depth dose curve
is fairly uniform, followed by a rapid
drop off
• Reference point must be placed on
the CAX @ the required depth
%DD (approx values for 10 x 10 field) (6MV)
100% - 1.5 cm
90% - 4 cm
50% - 15 cm
%DD (approx values for 10 x 10 field) (10MV)
100% - 2.5 cm
90% - 5.5 cm
50% - 18 cm
Advantages of Electrons
• Sharp dose fall off below
the surface
• Less absorption in bone
and cartilage
• Good cosmetic results
Disadvantages of Electrons
• Eyes; shields can cause scatter, bowing of the isocurves treats a larger area under the surface than
at the surface
Disadvantages regarding Dose
The Use of Bolus
Bolus, made of a tissue equivalent material, is often used in electron beam treatments for the following purposes:
●To increase the surface dose;
●To flatten out irregular surfaces;
●To reduce the electron beam penetration in some parts of the treatment field
Impact of Electrons
Electrons start interacting and depositing dose as soon as the skin surface is reached
At extended SSD:
• The lower % lines get wider
• The higher % lines (80-100%) get narrower and lose depth
Mycosis Fungoides
Mycosis fungoides, is the most common form of lymphoma.
It generally affects the skin, but may progress internally over time.
Is a class of non-Hodgkin’s lymphoma, which is a type of cancer of the immune system.
Total Skin Electron Therapy
Types of Electron Set Up
- Set Angles
Best Contact/Skin Apposition
• The field is setup using skin apposition
• This may mean variable gantry, collimator and floor angles each day
• Used for patients where
immobilisation/positioning may vary
• Used for patients where there is no concern of OAR or overlap of fields
Set Angles
• Used in regions where dose to OAR
require consideration
• Near previously irradiated fields or matching/close to current fields
• Requires accurate stable reproducible immobilisation
• Gantry, collimator and floor rotation are fixed as per the plan
What happens if best contact is not setup?
A setup that is not best contact may compromise coverage at a depth