Functions of anesthesia machine
First anesthesia delivery system
Ether soaked rag over patient’s mouth & nose
Dr. Crawford Long - March 30, 1842
Function of presure regulators
Reduces high & variable pressures in cylinders to intermediate pressure (45 PSIg)
O2 E Cylinder’s Volume & Pressure
Volume = 625L
Pressure = 2200 psig
N2O E Cylinder’s Volume & Pressure
Volume = 1590 L
Pressure = 725 psig
Air E Cylinder Volume & Pressure
Volume = 625 L
Pressure = 2000 psi
FiO2 calculation
FiO2 = O2 (L/min) + Air (L/min) (0.21)
Total Gas Flow (L/min)
Volatile agents’ MAC values
Iso = 1.2
Sevo = 1.7
Des = 6
Circle system components
Fresh Gas Inlet
Unidirectional Valves
Inspiratory Limb
Expiratory Limb
Gas Reservoir Bag (green)
Adjustable Pressure-Limiting (APL) Valve - on expiratory limb
CO2 Absorber
When and how to use the APL valve
Basic Ventilations modes
Spontaneous
Patient is breathing on their own without assistance
Assisted
Augmentation of tidal volume with patient making some respiratory effort (use PSV mode, or manually squeezing reservoir bag as patient inspires)
Controlled
Total control of respiration with patient making no respiratory effort (squeezing bag during bag/mask, or using ventilator after intubation)
How ascending bellows work
When to change CO2 absorber
What does noninterchangeable gas-specific connections to pipeline inlets (DISS) with pressure gauges, filter, and check valve do?
Prevent incorrect pipeline attachments; detect failure, depletion, or fluctuation
What is the purpose of pin index safety system (PISS) for cylinders with pressure gauges, and at least one oxygen cylinder?
Prevent incorrect cylinder attachments; provide backup gas supply; detect depletion
What is the purpose of low O2 pressure alarm?
Detect O2 supply failure at the common gas inlet
What is the purpose of minimum O2/N2O ratio controller device (hypoxic guard)?
Prevent delivery of less than 21% O2
What is the purpose of O2 failure safety device (shut-off or proportioning device)?
Prevent administration of N2O or other gases when the O2 supply fails
Why must O2 enter the common manifold downstream to other gases?
Prevent hypoxia in event of proximal gas leak
What is the purpose of O2 concentration monitor & alarm?
Prevent administration of hypoxic gas mixtures in event of a low-pressure system leak; precisely regulate O2 concentration
What is the purpose of automatically enabled essential alarms and monitors (eg, oxygen concentration)?
Prevent use of the machine without essential monitors
What is the purpose of the vaporizer interlock device?
Prevent simultaneous administration of more than one volatile agent
What is the purpose of capnography & anesthetic gas measurement as far as safety is concerned?
Guide ventilation; prevent anesthetic overdose; help reduce awareness
What is the purpose of O2 flush mechanism that does not pass through vaporizers?
Rapidly refill or flush the breathing circuit