Gas Distribution - High pressure machine components
High-pressure system receives gases from cylinders at high, variable pressures & reduces those pressures to lower, more constant pressure suitable for use in the machine
Gas distribution - Intermediate Pressure components
The intermediate pressure system receives gases from the pipeline inlet or the pressure regulator (cylinder)
Intermediate pressure = 37-55 psi
This is pipeline pressure or the pressure after cylinder regulator
Pressure in intermediate pressure system will drop to zero when master switch is turned off
Pipeline inlets & pressure gauges
Gas power outlet
Master switch (pneumatic component)
Oxygen flush valve
Oxygen pressure failure device (fail-safe) & alarm
Oxygen & nitrous oxide second-stage pressure regulators
Flowmeter valves
How O2 flush valve works & possible hazards associated with its use
Oxygen fail-safe device/alarm
-Device that shuts off or proportionally decreases all gases except oxygen or alarms when oxygen pressure has fallen to a dangerous level (usually <30 psig)
Oxygen fail-safe device/alarm: Datex Ohmeda
Oxygen fail-safe device/alarm: Drager
Secondary pressure regulators: function, gases used, final pressures, and purpose
Function - Reduce intermediate pressure to low pressure (20 psi O2, 38 psi N2O), to eliminate fluctuations in pressure supplied to the flow indicators caused by fluctuations in pipeline pressure
Low pressure machine components
Starts just after flow control valves
Hypoxic guard/proportioning systems (Link-25)
Flowmeter tubes
Vaporizers
Check valves (if present)
Common gas outlet
Hypoxic Guard System: Link-25 proprotioning system
If O2 supply pressure drops to 10 psi, in what order will the following things happen?
N2O flow stops
O2 fail safe alarm sounds
O2 fail-saffe device activates
Which of the following is not a mandatory safety feature of current anesthesia delivery systems?
A. Oxygen fail-safe device
B. Pulse Ox
C. Oxygen flowmeter at extreme right of flowmeter cluster
D. Vaporizer interlock device
E. All of the above are mandatory safety features
E. All of the above are mandatory safety features
If administering O2 and N2O on older Datex-Ohmeda machine, and lose pipeline O2 so switch to E cylinder without adjusting the flowmeters. What will happen to the floats in the flowmeter tubes?
A. both go to 0 and must be manually re-adjusted to their previous values
B. both floats go to 0 and rise back to their previous values
C. Both floats go to 0 and only the oxygen float rises back to its previous value, while N2O must be manually turned back on
B. both floats go to 0 and rise back to their previous values
Which of the following is not part of the low-pressure system of the anesthesia machine?
A. Vaporizer
B. Common gas outlet
C. Floats
D. Flowmeter tubes
E. O2 low pressure alarm
E. O2 low pressure alarm