fidelity of image
radiographic film quality
radiographic film quality are described by ____ and ____ factors
photographic and geometric factors
visibility of image
photographic factor
2 photographic factors
density and contrast
2 geometric factors
distortion and detail
darkness or blackness of image
radiographic density
radiographic density is evaluated by ____
optical density
measures the blackness of image
logarithmic function of original light intensity to the transmitted light intensity
optical density
⬆️ density = ___ OD
⬆️ density = ⬆️ OD
0 OD = ____ transmission
no transmission
Dmax = __ OD
4 OD
Dmin = 0.1 - ___
0.1 to 0.3
base density 0.11 - ___ OD
0.11 to 0.15 OD
fog density 0.15 - __ OD
0.15 to 0.18 OD
optimal density
0.25 to 2.5
used to measure the blackness
densitometer
density = direct (DaPaSaKaMa)
Destructive
Processing
Speed
kVp
mAs
density = inverse (OBCCSFG)
OID
Beam Restriction
Characteristics of Tissue (ATM)
Constructive
SID
Filtration
Grid
contrast = direct (OBGCC)
OID
Beam Restriction
Grid
Characteristics of Tissue
Contrast Media
contrast = inverse (3FiTE)
Fog
Filtration
Field Size
Thickness
Energy
shape distortion (PASS)
position
angulation
shape
size
sharpness of image and absence of blurring
detail
shadow of image or extra images on lateral side
penumbra
focal spot is in circular shape = __ penumbra
no penumbra