What is the typical kVP ranges for mobile device?
40-130 kVp
What is the typical mAs ranges for mobile device?
0.04-320 mAs
is the generator output at a high frequency or low frequency?
High frequency
is AEC available on some mobile untis?
Yes
What are 3 things to consider for the beam on mobile?
which side of the tube has a decrease of image density?
anode side
What is Anode Heel effect more pronounced with?
Optimal imaging requires that a grid be…
What is grid-cut off?
an unwanted absorption of x-rays via an x-ray grid
How should you angle grids for the direction of the rid lines?
angle in the direction of the grid lines (long axis), not against
What are three ways to prevent grid cut off?
How much off centered do you have to be mispositioned for transverse grid cut off?
1-1.5 inches
How to fix grid cut off from focal range?
images taken at distances greater or less than produce cutoff
What is off-level grid cutoff?
grid placed on a unstable surface such as the mattress of a bed and/or the grid to tilt “off level, causing grid cutoff
What is off-centered grid cutoff?
1-1.5” transverse off centering results in cutoff
What is off-focus grid cutoff?
images taken at distances greater or less than can produce cutoff
what is the upside-down grid cutoff?
inverting a focused grid causes pronounced grid cutoff
What grid cut offs cause overall loss of denstiy?
- Off center grid
What grid cut off causes reduced density on the lateral margins?
- Upside-down grid (most severe when grid is used upside-down)
how much kVp should be increased for every how many cm of body tissue?
Increase 2 kVp for every 1 cm of a body part
source to skin distance must not be less than what?
12”
When bringing mobile into a room wtih a patient supine, where is the base of the mobile?
postioned toward middle of bed
When bringing mobile into a room, with a patient seated, where should the base of mobile be?
parallel or 45 degrees toward the foot end of the bed
What are additional markers on mobile images?