High pressure system begins and ends at..
cylinder–> cylinder regulators
Intermediate-pressure system begins at and ends at …
pipeline and ends at flowmeter valves
The low-pressure system begins at the…. and ends at the …..
flowmeter tubes and ends at common gas outlet
What is SPDD?
Supply
Processing
Delivery
Disposal
PISS configurations:
Air:
Oxygen:
Nitrous Oxide:
Air: 1,5
Oxygen: 2,5
Nitrous Oxide: 3,5
Pressure change from cylinder to pipeline
1900 psi –> 45 psi
Pressure of cylinders:
Air:1900 psi
O2: 1900 psi
Nitrous: 745 psi
Max volumes of cylinders:
Air: 625L
Oxygen: 660L
Nitrous: 1590 L
How to calculate how many minutes a cylinder will last?
Cross multiply
Most delicate part of the cylinder?
cylinder valve
O2 pressure failure device:
alarm sounds when pipeline pressure falls below:
28-30 psi
What psi would stop the flow of nitrous oxide?
< 20 psi
will the oxygen pressure failsafe device detect a flowmeter leak? why?
no, it is upstream of the flowmeters
When using nitrous, the proportioning system never lets the O2 to drop below __%
25%
Hypoxia prevention safety device will not alarm in the following situations:
oxygen pipeline crossover
leaks distal to flowmeter valves
administration of third gas
defective mechanic or pneumatic components
To prevent hypoxic mixture in setting of a cracked flowmeter, the ____ should be positioned closest to the ______ (far ____).
O2
common gas outlet
Far right
What circumstance would create the risk of a hypoxic mixture when considering a crack in the flowmeters?
a crack in the O2 meter that is furthest to the right
FiO2 calculation:
FiO2 = Air flow rate x 21 +
O2 flow rate x 100 / total flow rate
What is the maximum O2 percentage a nasal canulla can deliver?
44% @ 6L
Compliance =
Volume / Pressure
The Tec 6 vaporizer heats the liquid anesthetic to __ C and pressurizes it to ____ _____.
39
2 atmospheres
Tec 6: A lower ambient pressure (higher altitude) requires a ___ setting on the dial.
higher
Tec 6: A higher ambient pressure (hyperbaric oxygen tank) requires a ____ setting on the dial.
lower
What device would be the first to detect an oxygen pipeline crossover?
oxygen analyzer