What is filtered back projection?
most commonly used algorithm - reverses the attenuation process and builds up the matrix where areas of high and low attenuation tend to reinforce eachother
What are the disadvantages of 1st generation scanners?
What are advantages and disadvantages of the third generation scanners?
Advantages
Disadvantages
What are two major ways the 4th generation scanner differs from the 3rd generation?
What determines slice thickness in single detector arrays?
Collimation
In single slice CT, what is the trade off between SNR and spatial resolution when increasing slice thickness?
For single slice detector arrays, the slice sensitivity profile is a consequence of what?
How do you calculate collimator pitch (single detector helical scanners)?
Collimator pitch = table movement (mm) per 360 gantry rotation ÷ collimator width (mm)
WHat does a pitch of 1 imply?
the table movement per gantry rotation is equal to the collimator width
What does a pitch < 1 imply
over scanning- higher image quality at higher dose to patient
What does a pitch >1 imply?
partial scanning - reduces scan time, patient motion, patient exposure, but increases volume averaging
What is binning?
electronic signals generated by adjacent detector elements are electronically summed
WHat determines slice thickness in multiple detector array scanners?
determined by the width of the detectors in the slice tickness dimension (z-axis) and not by collimation
How do you calculate detector pitch (multidetector scanners)?
detector pitch = table movement (mm) per 360 gantry rotation ÷ detector width (mm)
How do you calculate collimator pitch of MDA scanners?
Collimator pitch = detector pitch ÷ N (number of detector arrays)
What are some major advantages of MDA scanners?
What are disadvantages of MDA?
What are advantages of Helical CT scanning?
What are disadvantages of helical scanning?
In axial aquisition - what is the Multiple Scan Average Dose?
The does to tissue that includes the dose attibutable to scattered radiation eminating from all adjacent slices
How can you calculate/estimate the multiple scan average dose (MSAD)?
can be estimated with the CT dose Index (CTDI)
CTDI(mGy) = [reading (mGy) x length of chamber (mm)] ÷ beam width
Compare CTDIFDA to CTDI100
How do you calculatre radiation dose with helical acquisition?
Helical dose = CTDI dose ÷ collimator pitch
What two variable affect image resolution in CT?