1 - Why was automatic exposure control (AEC) developed?
a. To decrease the need for technique charts
b. To make the radiographer more of a “button pusher”
c. To determine the radiation exposure required to produce a quality image
d. To increase the productivity and the patient dose
SOLUTION: C
The purpose of AEC is to produce consistently quality images, requiring fewer repeats.
2 - Which of the following terms does not belong in the list?
a. Photodiode
b. Ion chamber
c. Ionization chamber
d. mAs readout
ANS: D - mAs readout
The mAs readout, while a component of AEC, is not another term for an AEC detector.
3 - Once a predetermined amount of radiation is transmitted through a patient to the AEC detector, which of the following occurs?
a. The x-ray exposure increases
b. The x-ray exposure decreases
c. The x-ray exposure terminates
d. A and B
ANS: C - The x-ray exposure terminates
Once a predetermined amount of radiation is transmitted through a patient to the AEC detector, the x-ray exposure terminates.
4 - Which automatic exposure device works by using a gas-filled chamber?
a. Phototimer
b. Diode sensor
c. Ionization chamber
d. Bucky proliferator
ANS: C - Ionization chamber
The ionization chamber contains air and, when exposed to x-radiation, creates an electrical charge due to ionization.
5 - Where is the sensor located in an automatic exposure device with a gas-filled chamber?
a. In front of the patient
b. Behind the Bucky tray
c. In front of the image receptor
d. On the tabletop
ANS: C - In front of the image receptor
The ionization chamber is an entrance-type device located immediately in front of the image receptor, measuring radiation after it has passed through the patient, tabletop, and grid.
6 - Which automatic exposure device works by converting x-ray photons first into light and then into an electronic signal?
a. Phototimer
b. Gas detector
c. Ionization chamber
d. Backup timer
ANS: A - Phototimer
The phototimer, using either a photodiode or photomultiplier tube, converts the absorbed radiation to light and then into an electrical signal that can be measured.
7 - Where is the sensor located in an automatic exposure device that converts light into an electrical signal?
a. In front of the patient
b. Behind the image receptor
c. In front of the image receptor
d. On the tabletop
ANS: B - Behind the image receptor
The phototimer AEC detector is typically located behind the image receptor. It is considered an exit-type device.
8 - Which exposure variable is controlled by the automatic exposure control device?
a. kVp
b. Focal spot size
c. Seconds
d. mA
ANS: C - Seconds
AEC only controls the length of the exposure (seconds). The radiographer must select the appropriate focal spot size, kVp, and mA (some units).
9 - What would the result be if a film-screen radiographic examination required an exposure time shorter than the minimum response time of the AEC sensors?
a. Image density would be optimal
b. Image density would be excessive
c. Response time has nothing to do with AEC
d. Image contrast would be high
ANS: B - Image density would be excessive
If the minimum response time is longer than what is needed to produce optimal density, the image will be overexposed, or have excessive density.
10 - What describes the shortest exposure time required for the AEC device to operate?
a. Backup time
b. Minimum response time
c. Master density time
d. A and B
e. A and C
ANS: B - Minimum response time
Minimum response time is the minimum amount of time that it takes for the exposure to be terminated.
11 - AEC eliminates the need for the radiographer to select the actual ____ the examination will require.
a. kVp
b. SID
c. Exposure time
d. Focal spot size
ANS: C - Exposure time
The role of AEC is to terminate the exposure at the right time. The radiographer must still select the SID, kVp, and focal spot size appropriate to the examination.
12 - When using AEC, what should the radiographer adjust to manipulate the contrast of the image?
a. kVp
b. Detector selection
c. Exposure time
d. Focal spot size
ANS: A - kVp
As with radiography in general, kVp is the controlling factor for image contrast.
13 - When using AEC with film-screen, what should the radiographer adjust to manipulate the overall density of the image?
a. kVp
b. Backup time
c. Focal spot size
d. Density control
ANS: D - Density control
In order to adjust the density of the film-screen image produced using AEC, the density control (i.e., +2, 0, –2) is the only variable to be adjusted.
14 - What is the appropriate setting for backup time/mAs?
a. 50% of the expected mAs
b. 100% of the expected mAs
c. 150% of the expected mAs
d. 400% of the expected mAs
ANS: C - 150% of the expected mAs
Backup time should be set at 150% to 200% of the expected exposure time (or mAs).
15 - What is the purpose of setting a backup time/mAs?
a. To prevent excessive exposure of the patient
b. To make certain the contrast is acceptable if the chosen kVp is too low
c. To make certain the density is acceptable if the patient moves
d. To prevent the selection of the wrong detector
ANS: A - To prevent excessive exposure of the patient
The backup time or mAs is the maximum amount of time that the x-ray exposure will stay on while using AEC. This is a safety mechanism to protect the patient from excessive exposure.
16 - Which of the following is a system that allows the radiographer to select a particular button on the control panel that represents an anatomic area and displays a preprogrammed set of exposure factors?
a. mAs readout
b. Anatomically programmed technique
c. Automatic exposure control
d. Phototiming
ANS: B - Anatomically programmed technique
Anatomically programmed technique is a system of preprogrammed technical factors that are displayed once a specific anatomic area and position (i.e., PA chest, oblique knee) is selected.
17 - When doing a film-screen study using AEC, what will happen to the density in the area of interest when changing from 70 to 90 kVp?
a. Density will increase
b. Density will decrease
c. Density will not change
ANS: C - Density will not change
Adjusting the kVp will not have any effect on the image density, because the detectors have terminated the exposure based on the amount of radiation transmitted. However, increasing the kVp will lower the image contrast.
18 - If the detector is directly exposed to the primary beam during a film-screen study, the area of interest will appear:
a. Overexposed
b. Underexposed
c. Correctly exposed
d. To have high contrast
ANS: B - Underexposed
When the detector is exposed directly to radiation, the exposure is terminated right away, and the anatomy of interest will be underexposed, or too light.
19 - In order to become familiar with manual technical factors (i.e., setting mAs and kVp and not using AEC), it is very helpful to pay attention to the:
a. Exposure time
b. Heat units
c. Density control
d. mAs readout
ANS: D - mAs readout
The mAs readout will briefly show the exact amount of mAs used for the previous exposure. It is important for the radiographer to pay attention to this readout so that there’s a greater familiarity with the mAs needed to produce optimal images.
20 - With modern x-ray equipment, the AEC system’s minimum response time is typically:
a. 1 ms
b. 10 ms
c. 50 ms
d. 100 ms
ANS: A - 1 ms
The minimum response time for most of today’s equipment is 1 ms.
21 - The most common cause for repeating images done with film-screen and AEC is inaccurate:
a. kVp selection
b. Backup mAs selection
c. Centering
d. Alignment of the tube and image receptor
ANS: C - Centering
Inaccurate centering is the primary reason for images to be repeated when using AEC with film-screen. If the anatomy of interest is not directly over the detector, the image will probably be less than optimal.
22 - The mAs readout:
a. Is of little value because it’s not necessary to know mAs when using AEC
b. Reports the actual mAs used for an exposure
c. Reports the patient dose for the entire imaging study
d. Allows the radiographer to better set the mAs for an AEC examination
ANS: B - Reports the actual mAs used for an exposure
The mAs readout shows the actual mAs used for the exposure just made. Awareness of this value helps in producing non-AEC images.
23 - With a CR system, when using AEC to image a patient who is having difficulty holding still, increasing the mA (if the unit allows) will:
a. Be a bad idea because the patient will be overexposed
b. Be a bad idea because the patient will be underexposed
c. Be a good idea because the exposure time will increase
d. Be a good idea because the exposure time will decrease
ANS: D - Be a good idea because the exposure time will decrease
Increasing the mA will result in a shorter exposure time, always helpful in reducing the impact of patient motion. The exposure to the patient and IR will remain the same.
24 - If the backup time/mAs is automatically set by the x-ray unit, the exposure should end when a maximum of has been reached.
a. 600 s
b. 600 kVp
c. 600 mA
d. 600 mAs
ANS: D - 600 mAs
It is required that x-ray units with AEC have a preset backup time or mAs which is 600 mAs.