a. directly proportional
b. inversely proportional
c. not related
a. directly proportional
The product of milliamperage and exposure time is directly proportional to the quantity of x-rays produced.
REF: p.148
a. increases
b. decreases
c. is not affected
a. increases
As mAs increases, the quantity of radiation reaching the IR increases.
REF: p.149
a. 62.5 mAs
b. 250 mAs
c. 125 mAs
d. 500 mAs
c. 125 mAs
mAs is calculated by multiplying mA by time in seconds. In this case, 250 × 0.5 = 125 mAs.
REF: p.149
a. Increases image brightness.
b. Decreases image brightness.
c. Does not directly affect image brightness.
c. Does not directly affect image brightness.
The level of mAs does not directly affect image brightness when using digital IRs. During computer processing, image brightness is maintained when the mAs is too low or too high.
REF: p.150
a. severely lower-than-needed mAs.
b. a little lower-than-needed mAs.
c. a little higher-than-needed mAs.
d. severely higher-than-needed mAs.
a. severely lower-than-needed mAs.
Severely lower-than-needed mAs produces an image with increased quantum noise, and higher-than-needed mAs exposes a patient to unnecessary radiation.
REF: p.152
a. image brightness
b. exposure indicator
c. accession number
d. optical density
b. exposure indicator
The numerical value that is displayed on the processed image to indicate the level of x-ray exposure received on the digital image receptor is the exposure indicator.
REF: p.150
a. mAs
b. kVp
c. SID
d. Central ray angle
a. mAs
With film-screen imaging, mAs directly controls the density produced on the image.
REF: p.151
a. increased; increased
b. increased; decreased
c. decreased; decreased
d. decreased; increased
b. increased; decreased
With film-screen imaging, when mAs is increased, density is increased; when mAs is decreased, density is decreased.
REF: p.151
a. insufficient; increase
b. insufficient; decrease
c. excessive; decrease
d. excessive; increase
a. insufficient; increase
When a film image is too light, it has insufficient density, and would require an increase in mAs by 2 on repeat to obtain a diagnostic image.
REF: p.151
a. 10%
b. 15%
c. 30%
d. 100%
c. 30%
A 25% to 30% adjustment in mAs is the minimum needed to demonstrate a visible change in density.
REF: p.151
a. 70 kVp
b. 80 kVp
c. 90 kVp
d. 100 kVp
d. 100 kVp
The higher the kVp, the greater amount of radiation reaching the image receptor and therefore, with film-screen imaging, the greater density.
REF: p.152
a. kVp
b. mA
c. Seconds
d. SID
a. kVp
kVp affects the exposure to the IR by altering the amount and penetrating ability of the x-ray beam.
REF: p.152
a. kVp
b. mA
c. Seconds
d. SID
a. kVp
kVp, by controlling the penetrating power of the x-ray beam, is considered the primary controller of film-screen radiographic contrast.
REF: p.152
a. increase
b. decrease
c. not affect
a. increase
Increasing the kVp increases the IR exposure and density produced on a film image, and decreasing the kVp decreases the IR exposure and density produced on a film image.
REF: p.153
a. density would be unchanged, and contrast would decrease.
b. density would increase, and contrast would decrease.
c. density would decrease, and contrast would increase.
d. there would be no change in either contrast or density.
c. density would decrease, and contrast would increase.
Decreasing kVp results in photons with lower penetrating power. Fewer photons are transmitted, resulting in decreased density, and even fewer penetrate the more dense parts, resulting in a film-screen image with increased contrast.
REF: p.154
a. lower, higher
b. higher, higher
c. higher, lower
d. It makes no difference.
c. higher, lower
Using a higher kVp and lower mAs is best, because the higher kVp provides more penetration, requiring less patient exposure.
REF: p.155
a. Double the mAs.
b. Halve the mAs.
c. Use one fourth of the mAs.
d. No change would be necessary.
a. Double the mAs.
Decreasing the kVp by 15% would require two times the mAs in order to maintain film-screen image density.
REF: p.154
a. 300 mA, 0.2 s
b. 400 mA, 0.2 s
c. 100 mA, 1 s
d. 100 mA, 0.001 s
c. 100 mA, 1 s
mA and exposure time, also expressed as their product (mAs), control radiographic density. 100 mA × 1 s (100 mAs) is the highest amount listed and will produce the greatest density.
REF: p.149
a. 300 mA, 0.5 s
b. 400 mA, 0.5 s
c. 100 mA, 0.5 s
d. 800 mA, 0.01 s
b. 400 mA, 0.5 s
mA and exposure time, also expressed as their product (mAs), control radiographic density. 400 mA × 0.5 s (200 mAs) is the highest amount listed and will produce the greatest density.
REF: p.149
a. reduce the mAs by 30%.
b. reduce the mAs by 50%.
c. increase the mAs by 30%.
d. increase the mAs by a factor of two.
b. reduce the mAs by 50%.
When the image density is so far off that the image must be repeated, the minimum amount of change needed is to either double or halve the mAs.
REF: p.151
a. 51 kVp and 60 mAs
b. 51 kVp and 30 mAs
c. 69 kVp and 30 mAs
d. 69 kVp and 15 mAs
d. 69 kVp and 15 mAs
In order to lower the contrast, a higher kVp is needed. However, because the density was appropriate, the mAs has to be reduced to maintain image density.
REF: p.157
a. less absorption and more transmission in anatomic tissues.
b. less variation in the x-ray intensities exiting the patient.
c. increase interactions from Compton scattering.
d. low-contrast image.
e. all of the above.
e. all of the above.
A high kVp results in less absorption and more transmission in anatomic tissues, which results in less variation in the x-ray intensities exiting the patient (lower subject contrast), producing a low-contrast image. High kVp results in a greater proportion of interactions from Compton scattering than x-ray absorption (photoelectric effect).
REF: p.157
a. angle of the anode used.
b. actual size of the filament used.
c. energy of electrons available for tube current.
d. distance the electrons travel from cathode to anode.
b. actual size of the filament used.
When selecting large or small focal spot at the console, what’s really being selected is the large or small cathode filament.
REF: p.158
a. magnification factor.
b. SID.
c. OID.
d. SOD.
b. SID.
The distance between the radiation source and the image receptor is the SID, or source to image receptor distance.
REF: p.158