PHYSICS Flashcards

(68 cards)

1
Q

pourquoi T2* plus sensible pour hémosidérose

A

T2*-weighted and other gradient-echo sequences are the most sensitive to magnetic susceptibility effects of ferritin and hemosiderin because of their lack of a 180-degree refocusing pulse.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

explique pourquoi dans mx polykystique rein tu peux voir kyste hyperechogene au lieu anechogene

A

Because of their small size, some of the cysts are hyperechoic (instead of anechoic) with parallel echogenic lines trailing from the far walls. This is due to reverberation artifact.

Reverberation artifact is caused by repeated reflections of intracystic sound energy between the near and far walls of the cyst, creating aberrant “copies” of the cyst wall in the tissue beyond.

–> dans le fond, transducer pense que chaque echo recoit une seule réflection –> donc les echos qui ont plus que une reflection cree une copie plus distale

This is due to a faulty assumption by the transducer that each echo it receives only encountered a single reflection. Because the distance of the reflector from the transducer is mapped based on a calculation contingent on the fixed speed of sound, echoes that are reflected more than once are artificially mapped deeper in tissue than they actually are. Multiple reflections create multiple lines deep to the cyst depending on the number of reflections each echo encounters (more reflections for a given echo equate to a greater mapped distance from the transducer).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

zone IRM

A

ZONE I : sans restriction (outside world)

Zone II refers to the interface between unrestricted zone I (i.e., the outside world) and restricted zone III (i.e., the control room immediately outside the scan room). Zone II is used for intake of patients scheduled to undergo MR imaging. It is a place where patient histories can be obtained and MR screening can be performed.

zone 3 : controle room, immédiatement dehors du scan

Zone IV is the maximum zone level and refers to the room that contains the magnet.

Locks or other barriers are required between zone II and zone III.
Unscreened individuals are not allowed to enter zone III or zone IV.

If a medical emergency occurs in zone IV, the patient should be relocated to a lower zone to permit safe resuscitative measures outside the magnetic environment.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

quelle parametre impact atténuation en CT

A

kVP

low kVp imaging is closer to the k-edge of iodine, resulting in increased attenuation at lower kVp within iodine-containing (i.e., enhanced) structure

Lower kVp imaging is associated with greater image noise, less radiation dose, and greater measured iodine attenuation than higher kVp imaging.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

paramètre de temps dans dual echo, quelle est plus long et pourquoi

A

When selecting echo times for dual-echo gradient-recalled echo imaging,
it is important that the in-phase echo time be longer than the opposed-phase echo
time.

This criterion permits microscopic fat to be differentiated from susceptibility
artifact.

If the in-phase echo time is shorter than the opposed-phase echo time,
microscopic fat and susceptibility artifact will both cause signal loss on the opposed phase images (i.e., microscopic fat because fat and water protons are in opposite alignment, and susceptibility artifact because of the opposed-phase echo time is
longer than the in-phase echo time [not desirable]).

(a 3T, 1 de plus; a 2T 2sec de plus)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Along what vector do the protons align in
conventional inversion recovery imaging immediately following the preparation
radiofrequency pulse (applied before the excitation pulse)?

A

With conventional spin echo–based inversion recovery imaging, a 180-degree preparation radiofrequency pulse is applied before a 90-degree excitation radiofrequency pulse

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

stepladder artefact ?

A

Reverberation artifact occurs when echoes are repeatedly reflected between two highly reflective interfaces. This results in multiple equidistantly spaced linear reflections which have a stepladder appearance.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

predicteur brulure peau en fluroscopie

A

Cumulative air kerma is an expression of radiation at a point, so it would best predict skin injury. Kerma area product is an expression of total energy deposition across the entire exposed skin so it would best predict risk of radiation induced cancer.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

sepration frequent gras et eau pour calculer chemical shift

A

fat/water resonant frequency
separation (3.5 ppm)

Shift = (Fat/water resonant frequency separation [in ppm])×(field strength [T] ×42MHz/T

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What organ is best used as an internal reference standard to determine the degree of signal loss within an adrenal nodule

A

RATE

The spleen has been shown to be the best internal control on chemical shift imaging when attempting to determine the degree of signal loss within the adrenal gland. This is because the spleen does not have internal fat content that might confound interpretation (such as is often present in the liver).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

artefact qui cause pseudorehaussement dans petite lesion renale

A

Erroneous beam hardening correction

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

double energy ct basé sur quel principe

A

DIFFÉRENCE K-EDGE ÉLÉMENT UNIQUE IMAGÉ À kVP DIFFÉRENT

Dual-energy CT is based on the principle of distinguishing elements that
have sufficiently unique K-edges by imaging them with two different kVp settings.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

V ou F, aucune image en IRM réflete purement T1 et T2 ?

A

V
Remember that no image reflects purely T1 information or purely T2 information. MR images are a combination of both, with “weighting” toward one or the other based on the desired image contrast.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

effet concentrated gadolinium-based
contrast material (GBCM)

A

indirectly shorten the T1 time of adjacent protons, but it also shortens T2 times.

This effect is not noticeable on most T1 weighted images because the T1 shortening effects dominate
gado = Gadolinium is a paramagnetic agent

However, when gadolinium becomes highly concentrated (e.g., within a dilated renal pelvis, or in the dependent urinary bladder), the T2-shortening effects take over and produce a signal void.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

risque cancer induit avec dose effective 10 msv chez adulte et ped

A

The risk of cancer induction in an adult patient who receives an effective dose of 10 mSv is estimated to be roughly 0.01% (1 in 1,000).

The risk likely is higher for young adults, children, infants, and fetuses. The risk to a newborn infant is estimated to be roughly 0.4% (1 in 250) per 10 mSv dose.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

c’est quoi le principe de doppler shift

A

The reflected ultrasound frequency always increases when the ultrasound
beam interacts with a structure moving toward it.

Likewise, the reflected ultrasound
frequency always decreases when the ultrasound beam interacts with a structure
moving away from it.

The ultrasound probe emits sound at a known frequency (f₀).
Moving structures in the body (e.g., red blood cells) reflect the waves back.
If the blood cells are moving toward the probe: The distance between the wavefronts is shortened.

Reflected frequency (fᵣ) is higher than the transmitted frequency.
If the blood cells are moving away from the probe: Wavefronts are stretched.
Reflected frequency is lower than f₀.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

relation frequence et wavelenght

A

As the frequency increases, the
wavelength decreases (i.e., becomes more compressed),

as the frequency
decreases, the wavelength increases (i.e., becomes more elongated).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

quelle parametre statistique varie avec prévalence maladie

A

VPP
VPN

the positive and negative predictive values depend on the prevalence of disease in the population; therefore, the PPV and NPV will be different if the test is applied to a population with different disease prevalence.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

quelle effet prédomine dans paramètre suivant

LONG TR, LONG TE
COURT TR, COURT TE

A

heavily T2 weighted (static fluid) = long TR, long TE, fast spin echobased sequence)
-> Fluid has a very long T2; therefore, on a heavily T2-weighted sequence, the signal from all other structures is suppressed while the signal from fluid is preserved.

T1 weighted (short TR, short TE,
gradient recalled echo-based sequence).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What is the primary function of statistical iterative reconstruction methods

A

to reduce image noise.

This is accomplished by matching a simulated CT projection to the acquired CT projection in an iterative (repeated) fashion to better model the image noise.
Reduced image noise can translate into radiation dose savings, but the effect is indirect

one can obtain noisier images
with less radiation exposure, and then “clean up” the noisy images with statistical
iterative reconstruction after they are acquired.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

frequence nyquist

A

represent The analog frequency that is equal to half the sampling frequency (fs)
fN = 0.5 × fS

Before analog data can be displayed in a digital format, it must first be converted
to digital data using an “analog-to-digital” converter (ADC). The sampling
frequency (fs) of the ADC must be at least twice that of the analog frequency to
preserve the integrity of the data.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

pour maximiser T1
tu veux augmenter ou diminuer TE ?
puis tu veux grand ou petit flip angle ?

A

PETIT T1, LARGE FLIP ANGLE

Short echo times (TE) minimize T2 (T2*) contrast because when the TE is short, there is insufficient time for spin–spin relaxation to occur, and so therefore all substances, regardless of their T2 time, will appear similar

Large flip angles promote T1 weighting by converting a large fraction of longitudinal
magnetization into transverse magnetization. This allows differences in T1 times to
become evident during spin–lattice relaxation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

catégorie de Susceptibility

A

Susceptibility reflects the tendency of an object to become magnetized when placed into an external magnetic field.

There are four major categories.
1. diamagnetique: paroi vessie
2 paramagnétique: gadolinium
3. superparamagnétique: particule fer oxide
4. ferromagnetique : arthroplastie metalique

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

comment diminuer flow voids

A

DIMINUER TEMPS TE

The 180° refocusing pulse occurs at TE/2.

A longer TE → longer time between 90° and 180° pulses.

More time = higher chance moving protons leave the slice (or new ones enter too late).

Therefore: longer TE increases flow voids.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
quel artefact cause dirty shadowing
reverberation artifact
26
effet frequence sonde sur penetrance
High-frequency transducers have less penetration and greater spatial resolution than do low-frequency transducers.
27
what is ADC
ADC is the slope of the plot ln(SI) at various b-values.
28
quelle parametre de lecho est independant de la direction du flux
Power Doppler imaging is direction independent, and color Doppler imaging is not
29
power vs color doppler
Power Doppler imaging is related to the amplitude of the echo (which is proportional to the number of flowing particles), while color Doppler imaging is related to the frequency of the echo (which contains information about the Doppler shift and, hence, direction of flow). Power Doppler imaging is direction independent, and color Doppler imaging is not AUTRE: Power Doppler imaging is more sensitive to slow flow than is color Doppler imaging unlike color Doppler imaging, it is not affected by aliasing (because unlike color Doppler imaging, power Doppler imaging is a magnitude-only, directionless expression).
30
quelle sequence utilise gradient bipolaire pour differentier spin rapide des spin stationnaire
Strong bipolar gradients are used in echo planar imaging and phase contrast imaging to differentiate moving from stationary spins. When bipolar gradients are activated in succession, they induce a phase shift in moving spins because the moving spins are affected differently by the two equal gradient pulses. Moving spins will have moved to a different location along the gradient by the time the second gradient pulse is activated, so the phase accumulated from the first gradient pulse is not cancelled by the second gradient pulse. This phase shift is related to the velocity of the moving spins and can be used to determine the speed of a moving structure (e.g., flow across a stenotic valve). Stationary spins (i.e., precessing protons in immobile tissue) will be equally affected by both gradients resulting in zero phase shift because the stationary spins will not have moved between the times of gradient activation.
31
formule calcule taille voxel
voxel size for a 2D MR image is calculated with the following formula: (FOV/phase-encoding steps) × (FOV/frequency-encoding steps) × (slice thickness).
32
comment diminuer artefact cinetique en IRM
Change the phase-encoding axis to “left-to-right The motion is visible only along the phase-encoding axis because the phase-encoding axis takes much more time to acquire than the frequency-encoding axis. By changing the phase-encoding axis to “left-to-right,” motion-related artifacts will also change direction and their effect on the organ will be minimized
33
paramètre qui diminue le plus dose
kVp : has an exponential relationship with dose, while mA , pitch and z-axis coverage have a linear relationship. Reducing the pitch will increase the dose to the patient, not decrease it.
34
effet FOV sur SNR et résolution spatial
augmente FOV --> voxel size increases, SNR increases and spatial resolution decreases.
35
effet augmenter bandwidth
Decreases chemical shift artifact Increases data acquisition speed Decreases SNR.
36
effet phase et frequence matrice encoding sur résolution spatial
increasing the frequency or phase encoding matrix --> decrease the voxel size --> resulting in less signal, less SNR, and greater spatial resolution. for 2D sequence, the formula for calculating voxel size is : (phase FOV/matrix) × (frequency FOV/matrix) × slice thickness.
37
comment augmenter temps acquisation dans séquence fast spin echo pulse
Reducing the echo train length will increase the image acquisition time for a multislice fast–spin-echo pulse sequence - augmenter temps répétiion - augmenter nombre etape encodage phase - augmenter nombre signal Time is proportional to : (TR × Np × NSA)/ETL where TR is the repetition time, Np is the number of phase-encoding steps NSA is the number of signal averages ETL is the echo train length (Echo Train Length (ETL) = the number of echoes collected within a single repetition time (TR) during a Fast Spin Echo (FSE)
38
quest qui cause le renforcement postérieure en echo
Posterior acoustic enhancement” (i.e., “increased through transmission”) is caused by two things: (1) lack of amplitude attenuation of the ultrasound beam as it passes through the cyst (2) attenuation correction artifact.
39
chemical shift artifact a lieu le long de quel axe
Chemical shift artifact only occurs along the frequency-encoding axis. ex: si tu vois lesion kystique avec portion superieur en hyposiganl et inferieur en hypersignal, lencodage a eu lien superieur a inférieur Chemical shift artifact of the first kind occurs along fat–water interfaces. The width of the artifact increases with increasing field strength and lower receiver bandwidth
40
effet gadolinium sur sequence en IRM
Gadolinium-containing contrast material primarily functions by INDIRECTEMENT shortening the T1 times of nearby tissue
41
STIR cest quelle genre technique
STIR is an inversion recovery technique. All tissues in the slice are subjected to a 180-degree inversion recovery RF pulse prior to the excitation RF pulse, with the timing of the 180-degree RF pulse being determined by the T1 time of fat. The length of time between the 180-degree inversion RF pulse and the excitation RF pulse is known as the inversion time (TI).
42
vitesse son en echo
speed of sound commonly assumed by ultrasound machines for all tissue types is 1,540 m/s.
43
limitation technique STIR
it is not specific to fat because substances with short T1 times similar to fat (e.g., hemorrhage, contrast material) also will be suppressed with this technique. This is why inversion recovery is not commonly used for fat suppression at T1-weighted imaging
44
explique principe flow voids dans artere
Signal generation with spin-echo MR imaging requires protons to be exposed to both a 90-degree excitation radio-frequency (RF) pulse and a 180-degree refocusing RF pulse. Protons that are not exposed to both RF pulses will not generate signal. Fast-flowing blood moving through the imaged section will only be exposed to a single RF pulse. This results in an intravascular signal void.
45
pk vaisseau hyperintense dans time of flight
Unsaturated residual longitudinal magnetization entering the image volume
46
for a conventional spin-echo–based sequence, when should the 180-degree refocusing pulse be applied
APRES TE/2msec du pusle d'excitayion 90 degre Spin-echo sequences are based on the principle of performing a 180-degree refocusing pulse TE/2 msec after the initial 90-degree excitation pulse.
47
conventional filtered back projection (FBP), what is the relationship of tube current to noise?
With filtered back projection, tube current is inversely proportional to noise. That is, increasing the mA by factor of 4 will yield half the noise (1/square root of 4). FBP = computer/software filter
48
what can changing the type of reconstruction algorithm/kernel affect ?
It can affect spatial resolution and noise depending on which algorithm/kernel is used. . Reconstruction algorithm/kernel does not affect radiation dose since it is applied after the study is already obtained.
49
How is dose length product (DLP) related to scan length?
DLP is obtained by multiplying the CTDIvol by the scan length DLP (Dose Length Product) CTDIvol × scan length DLP = Represents the total radiation output over the whole scan. CTDvol : A standardized measure of scanner radiation output, not the patient’s actual dose. It’s the dose per slice, averaged across a phantom, and adjusted for pitch.
50
what is and how to calculated effective dose
Effective dose gives a general population risk rather than patient-specific risk. Obtenu en multipliant DLP x facteur conversion (f). The conversion factor is obtained by Monte Carlo simulation, and the best estimates (f) factor should be size specific.
51
RELATION entre direction encodage frequence et direction phase encodage
frequency encoding direction is opposite the phase-encoding direction
52
what is the most likely cause of transient interruption of the contrast bolus from an injection in the right antecubital fossa
Increased flow from the IVC - respiration --> pression negative thorax --> augmente reflux sanguin VCI vers OD The RA/SVC dilate; contrast coming from the right arm via SVC may swirl and mix rather than stream efficiently forward, and the big IVC inflow dilutes it.
53
why does gadolinium have paramagnetic properties when placed in a magnetic field
Gadolinium has paramagnetic properties due to unpaired outer shell electrons. When in a magnetic field, gadolinium becomes temporarily magnetized. The interaction between the outer shell of electrons and adjacent hydrogen nuclei leads to the T1-shortening properties of gadolinium.
54
what is pitch and its effect on dose
Pitch is defined as the ratio of the table distance traveled in one full gantry rotation divided by the thickness of all the acquired slices. C onceptually, this means with a pitch of <1, the same area is irradiated more than once. As you increase the pitch, the same area is irradiated less, and hence, the radiation dose goes down.
55
What is an advantage of inversion recovery fat suppression over chemical fat suppression
STIR imaging does not require a high field strength magnet. Chemical fat suppression is specific to fat and will not suppress contrast enhancement. However, it is susceptible to field inhomogeneity, which can cause incomplete fat saturation. With a higher-strength magnet, there is greater separation of the fat and water peaks making fat suppression easier
56
Which measurements of CT radiation dose takes into account the anatomic region and tissue being scanned?
EFFECTIVE DOSE (measured in sieverts, Sv) takes into account the tissue being scanned and its relative radiosensitivity. It takes into account both sex and an age-averaged reference person. Effective dose was designed for estimating radiation exposure of entire populations. PEARL: CONVERSATION FACTOR va donc etre PLUS GRAND pour organe radiosensible (ex: body 18 vs 2 tete)
57
organe les plus radiosensible
HOMME : red bone marrow, colon, lung, and stomach. FEMME red bone marrow, colon, breast, lung, and stomach.
58
artefact causé par flux sanguin
PULSATION selon direction PHASE ENCODING : motion artifact that improperly assigns signal to alternative locations greatest in the phase-encoded direction.
59
quelle artefact peut etre utiliser pour montrer anevrisme tip basilaire
GHOSTING/PULSATION --> dans la direction de la phase encoding
60
quelle artefact en peripherie d'un scan d'une personne fkn obese
TRUNCATION - the scan reconstruction assumes that the beam attenuation is caused by tissue within the field of view. - The resulting reconstructed image will have high attenuation around the periphery of the image. POUR CORRIGER ARTEFACT - AUGMENTE FOV pour inclure tout les tissus - inconvenient : diminue résolution
61
ring artifact indique quoi et comment on le corrige
ring artifact indicating a defective detector element. Corrected by recalibration, cleaning, or replacement of the detector
62
chemical shift type 1 vu dans quelle situation
* Seen in the frequency-encoding but not phase-encoding direction
63
chemichal shift artefact pire dans 1.5 ou 3 tesla
DANS 3 TESLA : because the difference in precession frequency between water and fat protons is larger at greater field strengths
64
comment diminuer artefact susceptibilité
increasing receiver bandwidth increasing echo train length, applying corrective reconstruction algorithms to restore spatial fidelity TE COURT : diminished on spin-echo and short TE sequences compared to gradient-echo (GRE) and longer TE sequence A longer TE results in greater susceptibility because it allows more time for the T2* effect of iron to degrade the transverse magnetization vector needed to produce a signal signal loss on T2W (longer TE) is greater than on T1W (shorter TE) images.
65
artefact qui peut orienter vers une MAV en IRM
GHOSTING / PULSATION : direction de la phase d'encodage
66
IRM PARALLELE
Parallel MRI imaging (pMRI) is a class of image acquisition and reconstruction routines seeking to increase the amount of imaging data acquired within a given time. The primary benefit of pMRI is faster image acquisition. Other benefits include decreased motion artifact, reduced breath-hold time, shorter duration exams, and an increase in the number of series per exam. pMRI techniques are designed to generate a specific artifact on the acquisition side (aliasing in the image domain from the under-sampling of k-space). The primary artifacts associated with pMRI tend to be residual aliasing in the image, or inconsistencies introduced into the image by the reconstruction process. Because of under-sampling, pMRI artifacts tend to result in spatially varying degrees of noise and artifacts in which are more severe at the center of the image and with increasing R factors.
67
TE dans in and out phase pour 1.5 t VS 3
The echo times used are specific to 1.5 T field strength (approximately 2.1 and 4.2 msec). The echo times are halved at 3.0 T (approximately 1.1 and 2.2 msec)
68
explique moi le side lobe artefact en echo et comment diminuer
Un transducteur envoie un faisceau principal (main lobe) + plusieurs faisceaux secondaires plus faibles = side lobes. Un side lobe frappe une structure hyperéchogène (ex : paroi vésicale, calcul, aiguille). L’écho revient vers la sonde. L’appareil suppose à tort que cet écho provient du main lobe (parce qu’il place tous les retours sur l’axe principal). Résultat = une fausse structure ou un bright spot apparaît dans l’image, à un endroit où il n’y a rien. DIMINUER : changer langle