X-direction (horizontal) patient motion

Describe how a signogram works

What are the findings in a sinogram with:
horizontal lines of decreased count density due to decreased count density in those individual frames
What can a linogram be used to detect
Y-direction motion
How should flood field non-uniformity be evaluated?
daily flood fields
For a Tc-99m sestamibi tetrofosmin myocardial perfsuion scan, setting the pulse height analyzer to 160 keV with a +/- 10% energy window:
Decreased image count density
What is the standard energy window and peak for Tc-99m?
Why?
What is the standard energy window and peak for most modern detectors:
What is the standard energy window and peak for Tc-99m utilizing a CDZ detector?
energy resolution of 5-6%
For an ECG-gated post-stress myocardial perfusion SPECT scan acquired on a two-headed scintillation camera, rejection of 30% of cardiac cycles due to arrhythmic beat rejection will result in approaximately:
30% decrease in overall image count density
What defines definite infective endocarditis according to the Modified Duke Criteria?
What are the minor criteria used in the Modified Duke Criteria for IE?

What are the major criteria used in the Modified Duke Criteria for IE?

Explain the difference between cameras:
CZT-based cameras have a lower uniformity defect contrast trade-off
Describe the difference in crosstalk between:
Tc-99m sestamibi myocardial perfusion SPECT was performed in a 65 year old man. The VLA post-exercise tomograms below were processed without (A) and with (B) low-dose XR attenuation correction.
The decrease in anterior wall count density in the AC images (B) is attributable to?

Scatter from subdiaphragmatic radiotracer concentration
***In this case, non-AC images are normal for a male

Metal artifact reduction in the CT component of a PET/CT is advantageous in the assessment of PET myocardial perfusion imaging because such artifacts may:
impair PET relative uptake and quantitation
The AC quality control figure shown, displays a fused image: the CT used for the attenuation map overlaid with a red contour showing the outlines of the myocardium on SPECT. What AC-induced artifact is most likely created if the MPI SPECT/CT misregistration shown?

Apparent uptake decrease in the lateral wall

What will MPI SPECT/CT misregistration show?
When using a CT scan for PET attenuation correction, voxel A has a CT number of zero Hounsfield units (HU) and voxel B has a CT number of 400 HU. What best describes the linear attenuation coefficients, μ(A) and μ(B)?
μ(B) > μ(A)
How are pixel values in CT images expressed?
usually expressed in Hounsfield units (HU) on a normalized scale with respect to the values for water
What effect does attenuation correction in SPECT MPI have:
Define attenuation coefficient
characterizes how easily a volume of material can be penetrated by a beam of light, sound, particles or other energy or matter
What affects does AC SPECT MPI have in obese patients?